RelayPro Thoracic Stent-Graft System 28-N4-30-164-30URecalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of RelayPro Thoracic Stent-Graft System (127)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150584), received between Oct 15, 2025 and Oct 15, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Retraction Problem” (1 report).
2 FDA recall records name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 22, 2024Class IICompleted
The stent-graft inside the delivery system was the incorrect size.
Z-1383-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 20, 2022Class IIOpen, Classified
The product may be shorter than packaging indicates
Z-0998-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Retraction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2025Retraction Problem; Retraction Problem
"DELIVERY SYSTEM FAILURE: WHEN THE STENT GRAFT WAS DEPLOYED WITH THE CONTROLLER IN POSITION 2, THE DEPLOYMENT GRIP COULD NOT BE RETRACTED. THE STENT GRAFT COULD BE DEPLOYED BY ROTATING THE DEPLOYMENT GRIP, AND THE DEVICE WAS IMPLANTED IN THE DESIRED POSITION. ANOTHER RELAY PRO DEVICE WITH THE SAME CATALOG AND LOT NUMBER WAS USED ON THIS PATIENT, BUT THE SAME EVENT OCCURRED. THE STENT GRAFT COULD BE DEPLOYED BY ROTATING THE DEPLOYMENT GRIP, AND THE DEVICE WAS IMPLANTED IN THE DESIRED POSITION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. OPERATION TYPE: TEVAR BLOOD LOSS: NO BLOOD LOSS IMAGE AVAILABLE UPON REQUEST PRE-CASE PLAN AVAILABLE UPON REQUEST ADDITIONAL INFORMATION AVAILABLE UPON REQUEST. TWO SETS OF THE DELIVERY SYSTEMS WILL BE RETURNED. (TC# (B)(4)" PATIENT OUTCOME: "NO HEALTH DAMAGE TO THE PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23285442 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.
- Z-2160-2026 · Apr 22, 2026 · Open, ClassifiedRelayPro Thoracic Stent-Graft System (various sizes) Reference numbers: 28-N4-32-104-32U 28-N4-32-164-28U 28-N4-32-164-32U 28-N4-32-209-28U 28-N4-32-209-32U 28-N4-32-259-32U 28-N4-
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page