Skip to content

RelayPro Thoracic Stent-Graft System 28-N4-42-159-42URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150645), received between Jan 2, 2025 and Jan 2, 2025 — classified as 1 death.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2026Class IOpen, Classified

    Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

    Z-2160-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0998-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • Material Deformation1
  • Retraction Problem1

Patient problems

  • Cardiac Arrest1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathJan 2, 2025Retraction Problem; Material Deformation; Insufficient Device Problem Information; Retraction Problem; Material Deformation; Insufficient Device Problem Information

    "PATIENT HAD PREVIOUS TEVAR WITH CHRONIC TYPE B DISSECTION AND 7.2 CM ANEURYSM; HE RUPTURED AND WAS STABLE AT TIME OF IMPLANTATION. HIS PREVIOUS TEVAR EXTENDED TO 5 CM PROXIMAL TO CELIAC ARTERY; THE RUPTURE WAS AT THE DISTAL END OF THE EXISTING GRAFT. WE OPENED N4-42-159-42U REF#(B)(4) AND "BAREBACKED IT" INTO THE R CIA. DEVICE WENT UP WITH NO ISSUE; IN THE VISCERAL SEGMENT WE TOOK THE FLEXIBLE INNER SHEATH OUT OF HARD OUTER SHEATH "IN STEP 1" AND PLACED IT IN LANDING ZONE; WE MOVED TO STEP 2 AND DR (B)(6) ATTEMPTED TO TURN HANDLE COUNTER CLOCKWISE TO LINE UP D MARKER WITH GRAFT MARKER. DR (B)(6) HAS DEPLOYED OVER 50 RELAY PROS AND IS A TRAINER FOR TA. HE COMPLAINED THAT THE HANDLE KEPT RECOILING BACK TO ORIGINAL POSITION EVERY TIME HE TURNED IT; UNDER FLUORO, IT WAS CLEAR THAT THE D MARKER WASN'T MOVING, AND EVERY TIME HE TURNED THE HANDLE, THE NOSE CONE WOULD RECOIL DOWN TOWARD THE PATIENTS FEET. AFTER A FEW ATTEMPTS IT WAS CLEAR THE GRAFT WAS GETTING SCRUNCHED WITH EVERY ATTEMPTED TURN OF THE HANDLE. WE MOVED THE CONTROL PANEL TO STEP 1 IN AN ATTEMPT TO RECAPTURE THE GRAFT (STILL CONSTRAINED BY INNER SHEATH) AND WERE UNABLE TO (EVEN THOUGH WE HAVE DONE THIS BEFORE IN THE PAST).

    Patient problems: Cardiac Arrest; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21054799 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates