Skip to content

RelayPro Thoracic Stent-Graft System 28-N4-44-164-44URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576150652), received between Dec 16, 2025 and Dec 16, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2026Class IOpen, Classified

    Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

    Z-2160-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0998-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • No Apparent Adverse Event1
  • Patient Device Interaction Problem1

Patient problems

  • Chest Pain1
  • Insufficient Information1
  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 16, 2025Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Insufficient Device Problem Information; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Insufficient Device Problem Information; Patient Device Interaction Problem

    "DISTAL STENT GRAFT-INDUCED NEW ENTRY (DSINE): THE PATIENT WHO UNDERWENT TEVAR WITH THE RELAYPRO DEVICE CAME TO THE HOSPITAL WITH CHEST PAIN ON THE EVENING OF NOVEMBER 21. A CT SCAN REVEALED A DSINE IN THE VESSEL WHERE THE PERIPHERAL PART OF THE DEVICE WAS POSITIONED. THE PHYSICIAN DETERMINED TO PERFORM ADDITIONAL TEVAR ON A SEMI-URGENT BASIS ON NOVEMBER 25. A RELAYPRO NBS (28-N4-38-154-38U, 2312110207) WAS IMPLANTED PERIPHERALLY NEAR THE COMMON CELIOMESENTERIC TRUNK, AND THEN A RELAYPRO NBS (28-N4-46-164-42U, 2311150173) WAS IMPLANTED CENTRALLY. CONSIDERING THE POSSIBILITY OF A TYPE 3A ENDOLEAK, A ZENITH ALPHA CUFF (46 MM X 97 MM) WAS ADDITIONALLY IMPLANTED MORE CENTRALLY ON THE IMPLANTED RELAYPRO, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. PHYSICIAN'S COMMENT: THIS WAS MY FIRST RELAYPRO CASE, AND I WONDER IF THE SELECTED DIAMETER OF THE DEVICE WAS APPROPRIATE. OPERATION TYPE: TEVAR ANCILLARY DEVICE USED: 28-M4-46-200-46U (2310030017) BLOOD LOSS: AMOUNT IS UNKNOWN. IMAGE AVAILABLE: SEE ATTACHED NO PRE-CASE PLAN AVAILABLE DUE TO HOSPITAL POLICY NO ADDITIONAL INFORMATION AVAILABLE DUE TO HOSPITAL POLICY (TC#(B)(4)" PATIENT OUTCOME: "HEALTH DAMAGE WAS MINOR. THE PATIENT OUTCOME IS

    Patient problems: Chest Pain; Insufficient Information; Vascular Dissection

    MAUDE report 23823786 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page