Skip to content

RelayPro Thoracic Stent-Graft System 28-N4-42-259-42URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00843576150904), received between Oct 23, 2024 and Apr 25, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Endoleak” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2026Class IOpen, Classified

    Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

    Z-2160-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Endoleak1
  • Failure to Align1
  • Failure to Seal1
  • Insufficient Device Problem Information1
  • Material Invagination1

Patient problems

  • Clinical Endoleak1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 25, 2025Material Invagination; Failure to Align; Insufficient Device Problem Information; Material Invagination; Failure to Align; Insufficient Device Problem Information

    "PT READMITTED, DOES NOT APPEAR AS IF THEY WERE READMITTED DUE TO COMPLICATIONS OF GRAFT IMPLANT. ON FOLLOW UP CT IT APPEARS AS IF THERE IS A RETROFLEXION OF THE PROXIMAL RELAY PRO. PATIENT STILL HAS RUNOFF TO THE LOWER EXTREMITIES IT DR IS CONCERNED ABOUT LONG TERM DURABILITY." PATIENT OUTCOME: "I DO NOT HAVE THE FINAL OUTCOME OF WHAT DR DECIDED TO DO WITH PATIENT BUT THEY WERE STABLE AT TIME OF TALKING WITH THE DOCTOR."

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21912356 ↗
  • InjuryOct 23, 2024Failure to Seal; Device Endoleak; Failure to Seal; Device Endoleak

    "SUBJECT PRESENTED A FUSIFORM AORTIC ANEURYSM THAT WAS TREATED WITH TWO RELAY PRO NBS (28-N4-42-259-42U, LOT # 207140275 AND 28-N4-46-164-42U, LOT #2309200249) ON (B)(6)2024. SUBJECT WAS HOSPITALIZED POST PROCEDURE ON (B)(6) 2024 TO REDO TEVAR DUE TO TYPE 1A ENDOLEAK. PI HAS NOT BEEN ABLE TO DETERMINE IF THE ENDOLEAK IS RELATED TO THE PROCEDURE OR DEVICE." PATIENT OUTCOME: "ON (B)(6) 2024 THE ENDOLEAK HAS BEEN DETERMINED RESOLVED WITHOUT SEQUALAE."

    Patient problems: Clinical Endoleak

    MAUDE report 20518054 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page