Skip to content

RelayPro Thoracic Stent-Graft System 28-N4-36-154-32URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00843576150973), received between Nov 5, 2024 and Sep 15, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Separation” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2026Class IOpen, Classified

    Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

    Z-2160-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0998-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult or Delayed Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2025Difficult or Delayed Separation; Difficult or Delayed Separation

    "THE PATIENT NEEDED 2 TEVAR GRAFTS TO EXTEND DOWN THE THORACIC AORTA TO THE SMA ARTERY. THE CELIAC WAS CHRONICALLY OCCLUDED. THE PHYSICIAN USED THE RELAYPRO NBS FOR BOTH DEVICES WITHOUT A SHEATH. THE FIRST DEVICE WENT IN AND DEPLOYED WITHOUT ANY ISSUES. THE PHYSICIAN THEN INTRODUCED THE COMPLAINT DEVICE INTO THE PATIENT. THE DEVICE WAS ADVANCED SMOOTHLY THROUGH THE VASCULATURE AND THE GRAFT WAS THEN PUSHED OUT IN POSITION 1. NO ISSUES WERE NOTICED DURING THOSE STEPS. THE INNER SLEEVE WAS RETRACTED BACK IN POSITION 2 WITHOUT ANY ISSUES. DURING THE RETRACTION OF THE CLASPING MECHANISM OF STEP THREE I NOTICED THE SUPPORT WIRES MOVING BUT THE PROXIMAL CLASP ONLY MOVED 2-4MM. THE PHYSICIAN TRIED PUSHING THE CATHETER UP AND DOWN AS WELL AS ROTATING, BUT IT WOULD NOT RELEASE. WE GOT DOWN TO THE GREEN HYPOTUBE IN THE BACK AND TRIED CUTTING THAT LONGITUDINALLY TO SEE IF IT WOULD HELP AND IT DID NOT. EVENTUALLY WE WERE ABLE TO GET A SNARE AROUND THE CLASP WHILE MOVING THE NOSECONE UP AND THE 2 PIECES RELEASED. THE DEVICE WAS THEN REMOVED FROM THE BODY WITHOUT FURTHER ISSUES." PATIENT OUTCOME: "NO PATIENT CONSEQUENCE HAS BEEN REPORTED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23057525 ↗
  • MalfunctionNov 5, 2024

    DEPLOYING DEVICE IN THORACIC AORTA, THE INNER SHEATH JUMPED BACKWARDS CAUSING THE GRAFT TO "WIND-SOCK" AND DEPLOY LOWER THAN INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20609120 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page