RelayPro Thoracic Stent-Graft System 28-N4-42-159-38URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843576151000), received between Nov 20, 2024 and Mar 12, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2026Class IOpen, Classified

    Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

    Z-2160-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation2
  • Material Invagination2
  • No Apparent Adverse Event2
  • Detachment of Device or Device Component1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 12, 2026Material Invagination; Material Deformation; No Apparent Adverse Event

    "ON FOLLOW UP CT THE DEVICE HAS INVAGINATED AT THE PROXIMAL EDGE. PRE AND POST OP IMAGING WILL BE PROVIDED." PATIENT OUTCOME: "THIS WAS FOUND ON ROUTINE F/U CT, PATIENT IS STABLE. NO ADDITIONAL TREATMENT HAS BEEN DONE YET."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24584378 ↗
  • InjuryFeb 14, 2026Material Invagination; Material Deformation; No Apparent Adverse Event

    "ON FOLLOW UP CT, THE DEVICE HAS INVAGINATED AND IS NO LONGER SEALING THE DISSECTION. PRE AND POST OP IMAGING WILL BE PROVIDED." PATIENT OUTCOME: "THIS WAS FOUND ON ROUTING F/U CT, PATIENT IS STABLE. NO ADDITIONAL TREATMENT HAS BEEN DONE YET."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24340442 ↗
  • MalfunctionNov 20, 2024Retraction Problem; Detachment of Device or Device Component; Retraction Problem; Detachment of Device or Device Component

    "DURING POST STENT DEPLOYMENT IN ZONE ¾ OF THE THORACIC AORTA THE PATIENT LOOKED TO HAVE A PRIOR POST ELEPHANT TRUNK. POST DEPLOYMENT OF RELAY PRO STENT WAS SUCCESSFUL IN ZONE ¾ , BUT RECAPTURING THE TIP (NOSE CONE) SEEMED TO BE DIFFICULT DUE TO THE ANGLE OF ANATOMY AND POST SURGICAL INTERVENTION FROM PRIOR REPAIR. SEVERAL ATTEMPTS WHERE MADE TO CAPTURE THE TIP (NOSE CONE) AND AFTER A FEW TRIES WE WERE ABLE TO SUCCESSFULLY MARRY THE NOSE CONE TO THE OUTER SHEATH PER IFU. NO SHEATH WAS USED DURING THE CASE, THE VESSELS WHERE DILATED USING TRADITIONAL DILATION TECHNIQUE USING MULTIPLE DILATORS. THE DEVICE WAS PULLED DOWN TO THE LEVEL OF THE LEFT ACCESS ARTERIOTOMY (LCFA 9.4MM) AND AT THAT POINT WAS READY TO BE REMOVED OVER A STIFF WIRE. IN THE PROCESS OF THIS ACTION, AS THE DEVICE WAS BEING PULLED THROUGH THE ARTERIOTOMY SITE; THE TIP (NOSE CONE) WAS DETACHED FROM THE MAIN DEVICE AND STAYED IN THE ARTERY." PATIENT OUTCOME: "PATIENT NEEDED CUT DOWN TO REMOVE THE TIP (NOSE CONE) FROM THE ARTERY."

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20738050 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2