Skip to content

RelayPro Thoracic Stent-Graft System 28-N4-34-209-30URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of RelayPro Thoracic Stent-Graft System (127)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
1
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843576151062), received between Sep 29, 2025 and Jan 2, 2026 — classified as 1 death, 2 injuries.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (P200045); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 22, 2026Class IOpen, Classified

    Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

    Z-2160-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 20, 2022Class IIOpen, Classified

    The product may be shorter than packaging indicates

    Z-0998-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Patient Device Interaction Problem1
  • Patient-Device Incompatibility1
  • Separation Failure1

Patient problems

  • Stroke/CVA2
  • Hemorrhage/Blood Loss/Bleeding1
  • Paraplegia1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 2, 2026Patient-Device Incompatibility

    "SUBJECT PRESENTED AN INTRAMURAL HEMATOMA (IMH) AND WAS IMPLANTED BY A RELAYPRO NBS ON (B)(6) 2024. THE PATIENT WAS ADMITTED ON (B)(6) 2024 WITH POST-OPERATIVE STROKE (CVA). INVESTIGATOR DEEMED THIS EVENT RELATED TO PROCEDURE AND DEVICE. SUBJECT UNDERWENT THROMBECTOMY WITH STENT PLACEMENT AND WAS DISCHARGE ON (B)(6) 2024. PENDING SOURCE DOCUMENTATION AND IMAGING FOR STROKE AS WELL AS CONFIRMATION OF DEVICES USED." PATIENT OUTCOME: "PATIENT WAS DISCHARGE ON (B)(6) 2024 AND THE EVENT WAS RESOLVED WITHOUT SEQUALE."

    Patient problems: Stroke/CVA

    MAUDE report 23944208 ↗
  • InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    "PARAPLEGIA: THE RELAY PRO STENT GRAFT WAS IMPLANTED TO TREAT A DESCENDING AORTIC ANEURYSM THAT WAS LOCATED AROUND THE T9-T10 SEGMENT OF THE THORACIC SPINAL CORD. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY ENDOLEAK. HOWEVER, AFTER THE PROCEDURE, THE PATIENT COULD NOT MOVE BOTH LEGS, AND PARAPLEGIA IN THE LOWER HALF OF THE BODY WAS IDENTIFIED. SINCE THE PATIENT COULD MOVE BOTH ARMS AND HAD A SENSE OF TOUCH IN THE LEGS, THE PATIENT WAS TRANSFERRED TO THE ICU FOR OBSERVATION. PHYSICIAN'S COMMENT: THE PROCEDURE WAS PERFORMED UNDER THE UNDERSTANDING OF BOTH THE PHYSICIAN AND THE PATIENT THAT IMPLANTING A STENT GRAFT WOULD COVER THE ADAMKIEWICZ ARTERY, WHICH IS LOCATED AT THE DISTAL END OF THE AORTIC ANEURYSM. THIS EVENT OCCURRED DUE TO THE IMPLANTATION OF A STENT GRAFT, BUT NOT BECAUSE THE STENT GRAFT WAS A RELAY PRO. PHYSICIAN'S COMMENT ON THE CAUSAL RELATIONSHIP: THERE WAS NO DIRECT CAUSAL RELATIONSHIP BETWEEN THE DEVICE AND THIS EVENT. (THIS EVENT WOULD HAVE OCCURRED EVEN IF IT WERE NOT A RELAY PRO.) OPERATION TYPE: TEVAR. NO BLOOD LOSS. ANCILLARY DEVICE USED: 28-N4-32-209-32U (LOT: 2410020354). NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY. NO PRE-CASE PLAN AVAILABLE DUE TO HO

    Patient problems: Paraplegia

    MAUDE report 23645248 ↗
  • DeathSep 29, 2025Separation Failure; Separation Failure

    "AS STATED ABOVE, PREVIOUS 30X32X 100 THORAFLEX IMPLANTED WEEK PRIOR (ON (B)(6)) WITH THE PLAN TO EXTEND WITH A TEVAR DOWN TO JUST PROXIMAL OF CELIAC. - CHRONIC DISSECTION. AFTER DISCUSSION OF CASE PLAN, WE CHOSE 34/30 X 200 TAPERED NBS. AS THE PROCEDURE COMMENCED, ALL WAS NORMAL THROUGHOUT EACH STEP OF DEPLOYMENT UNTIL STEP 3 WHEN, ALTHOUGH THE RELEASE MECHANISM DID TRAVERSE CAUDALLY (ON THE STERILE FIELD) WITH NO ISSUE AND AS PLANNED, & WE MOVED TO STEP 4 IN AN EFFORT TO REMOVE THE PUSHER ROD (PULL BACK) WE SAW THAT THE ROD WOULD NOT TRAVERSE OVER THE WIRE (WHILE ALSO WATCHING ON FLOURO) TO SEE THAT THE STENT GRAFT WAS TRAVERSING (MOVING) WITH THE ROD ITSELF ON GENTLE MOVEMENT. IT WAS THEN THAT WE DECIPHERED THAT THE CLASP WAS NOT RELEASED FROM THE STENT GRAFT. AFTER WORKING UNDER MY DIRECTION TO TROUBLESHOOT WITH FIRST STEP TO GENTLY MANEUVER THE PUSHER ROD UP/DOWN AND TWIST SIDE TO SIDE FOLLOWED BY CUTTING THE HYPO TUBE & MOVING THE (GREEN) TUBE IN A CAUDAL DIRECTION, THERE WAS NO CHANGE. I THEN REACHED OUT TO R&D TEAM (B)(6) TO PROVIDE DIRECTION IN WHICH THEY CONNECTED WITH ME VIA TEAMS CALL WHEREBY I WAS ABLE TO PROVIDE LIVE FEED OF OR AND WORK TO TROUBLESHOOT. AFTER 4 HRS OR

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Stroke/CVA

    MAUDE report 23169187 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P200045. One submission often covers many models, so check the product and lot details in each record.

  • Z-1383-2024 · Feb 22, 2024 · Completed
    RELAY PRO Thoracic Stent-Graft System, containing one endovascular stent-graft system, IFU, and patient tracking card, sterile: (a) Part number 28-M4-34-100-34U; (b) Part number 2

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page