TREO Abdominal Stent-Graft System 28-L2-20-100U (UDI 00843576152090)Recalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of TREO Abdominal Stent-Graft System (73)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576152090), received between Apr 23, 2026 and Apr 23, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Device Endoleak” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 3, 2023Class IIOpen, Classified
Potential for the incorrect size stent-graft than the printed carton label.
Z-1689-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Endoleak1
Patient problems
- Clinical Endoleak1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2026Device Endoleak
I WAS MADE AWARE LAST NIGHT OF AN EXPLANT OF A TREO DEVICE, PHYSICIAN STATED THERE WAS TYPE 3 ENDOLEAK AND THE DECISION WAS MADE TO EXPLANT THE DEVICE. NO FURTHER DETAILS WERE PROVIDED. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2026 FROM THE TERUMO AORTIC SITE REPRESENTATIVE: AT THE TIME OF IMPLANT BILATERAL RENAL ACCESSORIES WERE EMBOLIZED. POST IMPLANT ANGIOGRAM REVEALED SUCCESSFUL EXCLUSION OF ANEURYSM WITHOUT EVIDENCE OF ENDOLEAK. A CT PERFORMED ON (B)(6) 2021 SHOWED A POSSIBLE TYPE 2 ENDOLEAK "LIKELY DUE TO A NON-EMBOLIZED ACCESSORY LEFT RENAL", SAC WAS STABLE. HE HAD MULTIPLE INTERVENTIONS FOR A TYPE 2 ENDOLEAK WHICH LED TO A LAPAROSCOPIC LIGATION OF IMA AND MULTIPLE LUMBARS ON (B)(6) 2024. HE HAD A CV MRA ON (B)(6) 2025 THAT SHOWED SAC 7.2 CM, THE REPORT ALSO STATED "RIGHT COMMON ILIAC- DISTAL SEGMENT IS ECTATIC, THE ILIAC LIMB OF AORTOILIAC STENT GRAFT EXTENDS INTO THE COMMON ILIAC ARTERY. INCOMPLETE APPOSITION OF THE DISTAL END OF THE STENT GRAFT." EXPLANT PERFORMED 03/28/2026. ON THE OP REPORT IT STATES 'SIGNIFICANT BLEEDING WAS NOTED FROM A DEFECT IN THE GRAFT FROM THE POSTERIOR MAIN BODY. PATIENT OUTCOME: "NO INFO PROVIDED ON PATIENT. ADDITIONAL INFORMATION RECEIVED ON
Patient problems: Clinical Endoleak
MAUDE report 24967241 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page