TREO Abdominal Stent-Graft System 28-L2-15-160URecalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of TREO Abdominal Stent-Graft System (73)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843576152281), received between Mar 19, 2025 and May 28, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Endoleak” (3 reports).
1 other recall cover devices cleared under the same FDA submission (P190015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Endoleak3
- Failure to Seal3
Patient problems
- Clinical Endoleak3
- Aneurysm expansion1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 28, 2025Failure to Seal; Device Endoleak; Failure to Seal; Device Endoleak
"THE CORE LAB SENT NOTIFICATION OF A SIGNIFICANT FINDING ON (B)(6) 2025 FOR SUBJECT (B)(6) REGARDING A >5MM INCREASE SAC ENLARGEMENT FOR 2-YEAR IMAGING DATED (B)(6) 2024 AS COMPARED TO THE 30-DAY IMAGING. THE 2-YEAR CT WAS NON-CONSTRAST, THEREFORE THE CORE LAB ALSO USED THE RECENT 3-YEAR CT WITH/WITHOUT CONTRAST FOR COMPARISON OF CONTINUED GROWTH AS NOTED ON THE FINDING. PENDING SITE DETAILS" PATIENT OUTCOME: "PENDING SITE DETAILS".
Patient problems: Aneurysm expansion; Clinical Endoleak
MAUDE report 22100140 ↗InjuryMar 19, 2025Failure to Seal; Device Endoleak; Failure to Seal; Device Endoleak
"I WAS NOTIFIED THAT MS. (B)(6) CAME IN ON (B)(6) 2025 EARLY IN THE MORNING WITH A RUPTURED ILIAC LIMB. UPON OPENING THE PATIENT, THE DOCTORS WERE ABLE TO CONTROL THE BLEEDING AND REPLACED THE STENT GRAFT LIMBS WITH DACRON GRAFTS. IT WAS NOTED THAT IT LOOKED TO BE THAT THE CA+ HAD ERODED INTO THE STENT GRAFT LIMBS THUS CAUSING THE LEAK. PATIENT SEEMS TO BE DOING FINE." PATIENT OUTCOME: "SHE SEEMS TO BE DOING FINE."
Patient problems: Clinical Endoleak
MAUDE report 21646131 ↗InjuryMar 19, 2025Failure to Seal; Device Endoleak; Failure to Seal; Device Endoleak
"I WAS NOTIFIED THAT MS. (B)(6) CAME IN ON (B)(6)2025 EARLY IN THE MORNING WITH A RUPTURED ILIAC LIMB. UPON OPENING THE PATIENT, THE DOCTORS WERE ABLE TO CONTROL THE BLEEDING AND REPLACED THE STENT GRAFT LIMBS WITH DACRON GRAFTS. IT WAS NOTED THAT IT LOOKED TO BE THAT THE CA+ HAD ERODED INTO THE STENT GRAFT LIMBS THUS CAUSING THE LEAK. PATIENT SEEMS TO BE DOING FINE." PATIENT OUTCOME: "SHE SEEMS TO BE DOING FINE."
Patient problems: Clinical Endoleak
MAUDE report 21646130 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P190015. One submission often covers many models, so check the product and lot details in each record.
- Z-1689-2023 · May 3, 2023 · Open, ClassifiedThe TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graf
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page