Skip to content

TREO Abdominal Stent-Graft System 28-B2-26-080U (UDI 00843576152359)Recalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of TREO Abdominal Stent-Graft System (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576152359), received between Aug 18, 2025 and Aug 18, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Device Endoleak” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 3, 2023Class IIOpen, Classified

    Potential for the incorrect size stent-graft than the printed carton label.

    Z-1689-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Endoleak1
  • Failure to Seal1
  • Obstruction of Flow1

Patient problems

  • Aneurysm expansion1
  • Clinical Endoleak1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 18, 2025Obstruction of Flow; Failure to Seal; Device Endoleak; Obstruction of Flow; Failure to Seal; Device Endoleak

    "TREO BIFURCATE WAS OPENED AND PREPPED AS NORMAL. THE DEVICE WAS ADVANCED, POSITIONED AND DEPLOYED AS NORMAL. CUFF WAS USED TO HELP WITH TYPE 1A ENDOLEAK. A SECOND 20X120 LIMB WAS INTRODUCED INTO THE RCIA AS THE FIRST 20X120 LIMB LANDED SHORT DUE TO TORTUOSITY. THE CASE COMPLETED AS NORMAL AND A TERUMO AORTIC BALLOON WAS USED TO ON PROXIMAL SEAL AND ALONG ALL LIMBS. FINAL ANGIO SHOWED NO EVIDENCE OF ANY LEAKS AND GOOD PERFUSION THROUGHOUT RENALS AND HYPOGASTRICS." PATIENT OUTCOME: "PHYSICIAN CONTACTED REP ON (B)(6) 2025 AS PATIENT WAS COMING INTO ER WITH LIMB THROMBOSIS AND SHORT RENAL ARTERY OCCLUSION. DR (B)(6) STATED HE WAS PLANNING TO ATTEMPT TO THROMBECTOMIZE THE LIMB WITH PENUMBRA AND RELINE WITH A STENT OR GORE."

    Patient problems: Aneurysm expansion; Clinical Endoleak; Thrombosis/Thrombus

    MAUDE report 22832896 ↗

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page