TREO Abdominal Stent-Graft System 28-B2-30-080U (UDI 00843576152373)Recalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of TREO Abdominal Stent-Graft System (73)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00843576152373), received between Mar 11, 2025 and Oct 31, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Endoleak” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 3, 2023Class IIOpen, Classified
Potential for the incorrect size stent-graft than the printed carton label.
Z-1689-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Endoleak1
- Difficult or Delayed Separation1
- Mechanical Jam1
Patient problems
- Clinical Endoleak1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 31, 2025Device Endoleak; Device Endoleak
"POSSIBLE HOLE IN THE FABRIC FROM INITIAL IMPLANT. PLANNING TO ADD LIMBS BILATERALLY TO RAISE BIFURCATION. UPDATE ON 10/07/2025 FROM (B)(6) READER: THE PHYSICIAN BROUGHT THE PATIENT BACK ON SATURDAY (B)(6) 2025 AND IT WAS IN FACT TYPE 4, HOLE IN GRAFT OR SOME SORT OF TEAR. WE USED (2) 15X80 LIMBS TO RAISE THE BIFURCATION AND IT SEALED." PATIENT OUTCOME: "I SPOKE TO THE PHYSICIAN YESTERDAY (B)(6) 2025) AND HE IS DOING GREAT. BACK PAIN HAS BEEN RESOLVED AND HE WAS DISCHARGED ON SUNDAY (B)(6) 2025)".
Patient problems: Clinical Endoleak
MAUDE report 23441555 ↗MalfunctionMar 11, 2025Mechanical Jam; Difficult or Delayed Separation; Mechanical Jam; Difficult or Delayed Separation
"THE PHYSICIAN IMPLANTED THE DEVICE UNDER NORMAL CIRCUMSTANCES THROUGH A 20FR COOK SHEATH. HE DEPLOYED THE DEVICE AND WENT TO RELEASE THE PROXIMAL CLASP BY ENGAGING THE GREY GRIP RELEASE HOLDER. THE GREY GRIP WOULD NOT MOVE. HE PUSHED VERY HARD FORWARD AND IT WOULD STILL NOT RELEASE. I MADE SURE HE WAS NOT HINDERING THE BLACK RELEASE IN ANY WAY. HE THEN TRIED MOVING IT WITH LARGE CLAMPS BUT THAT ALSO DID NOT RELEASE IT. HE THEN BROKE OFF SIDES OF THE BLACK GRIP WITH HEMASTATS AND WAS ABLE TO PULL THE BLACK GRIP BACK AND RELEASE THE PROXIMAL CLASP." PATIENT OUTCOME: "THE DEVICE WAS THEN FULLY RELEASED AND THE PATIENT IS DOING FINE AT LAST CHECK."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21573209 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page