Skip to content

TREO Abdominal Stent-Graft System 28-B2-30-100U (UDI 00843576152458)Recalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of TREO Abdominal Stent-Graft System (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576152458), received between Dec 16, 2024 and Dec 16, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Obstruction of Flow” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 3, 2023Class IIOpen, Classified

    Potential for the incorrect size stent-graft than the printed carton label.

    Z-1689-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Obstruction of Flow1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Thromboembolism1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2024Obstruction of Flow; Patient Device Interaction Problem; Obstruction of Flow; Patient Device Interaction Problem

    "FAILURE DUE TO THE TECHNIQUE: THE ORIGINAL PLAN WAS TO PLACE THE STENT-GRAFT JUST BELOW THE RENAL ARTERIES. HOWEVER, THE AORTA JUST BELOW THE RENAL ARTERIES WAS CURVED, AND THE STENT-GRAFT BECAME CAUGHT IN THE CURVE DURING DEPLOYMENT. THE PHYSICIAN THOUGHT THAT THE STENT-GRAFT WAS PLACED JUST BELOW THE RENAL ARTERIES, BUT IT TURNED OUT THAT THE STENT-GRAFT WAS COVERING THE RENAL ARTERIES. AS BOTH RENAL ARTERIES MUST REMAIN PATENT, CANNULATION OF THE RENAL ARTERIES AND PERFORMANCE OF SNORKEL TECHNIQUE WERE PLANNED. CANNULATION OF THE RENAL ARTERIES WAS SUCCESSFULLY PERFORMED, THE CORONARY STENTS WERE THEN PLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. FINAL ANGIOGRAPHY REVEALED THAT THE RENAL ARTERIES WERE VISIBLE AND NO ENDOLEAKS WERE OBSERVED. OPERATION TYPE: EVAR BLOOD LOSS: UNKNOWN. NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY NO PRE-CASE PLAN AVAILABLE DUE TO HOSPITAL POLICY NO ADDITIONAL INFORMATION AVAILABLE DUE TO HOSPITAL POLICY DELIVERY SYSTEM WAS DISCARDED BY USER . (TC#BM 241102309)". PATIENT OUTCOME: "IT IS UNKNOWN WHETHER THERE WAS ANY HEALTH HARM TO THE PATIENT'S HEALTH."

    Patient problems: No Clinical Signs, Symptoms or Conditions; Thromboembolism

    MAUDE report 20943883 ↗

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates