TREO Abdominal Stent-Graft System 28-B2-30-100U (UDI 00843576152458)Recalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of TREO Abdominal Stent-Graft System (73)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576152458), received between Dec 16, 2024 and Dec 16, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Obstruction of Flow” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 3, 2023Class IIOpen, Classified
Potential for the incorrect size stent-graft than the printed carton label.
Z-1689-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Obstruction of Flow1
- Patient Device Interaction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Thromboembolism1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 16, 2024Obstruction of Flow; Patient Device Interaction Problem; Obstruction of Flow; Patient Device Interaction Problem
"FAILURE DUE TO THE TECHNIQUE: THE ORIGINAL PLAN WAS TO PLACE THE STENT-GRAFT JUST BELOW THE RENAL ARTERIES. HOWEVER, THE AORTA JUST BELOW THE RENAL ARTERIES WAS CURVED, AND THE STENT-GRAFT BECAME CAUGHT IN THE CURVE DURING DEPLOYMENT. THE PHYSICIAN THOUGHT THAT THE STENT-GRAFT WAS PLACED JUST BELOW THE RENAL ARTERIES, BUT IT TURNED OUT THAT THE STENT-GRAFT WAS COVERING THE RENAL ARTERIES. AS BOTH RENAL ARTERIES MUST REMAIN PATENT, CANNULATION OF THE RENAL ARTERIES AND PERFORMANCE OF SNORKEL TECHNIQUE WERE PLANNED. CANNULATION OF THE RENAL ARTERIES WAS SUCCESSFULLY PERFORMED, THE CORONARY STENTS WERE THEN PLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. FINAL ANGIOGRAPHY REVEALED THAT THE RENAL ARTERIES WERE VISIBLE AND NO ENDOLEAKS WERE OBSERVED. OPERATION TYPE: EVAR BLOOD LOSS: UNKNOWN. NO IMAGE AVAILABLE DUE TO HOSPITAL POLICY NO PRE-CASE PLAN AVAILABLE DUE TO HOSPITAL POLICY NO ADDITIONAL INFORMATION AVAILABLE DUE TO HOSPITAL POLICY DELIVERY SYSTEM WAS DISCARDED BY USER . (TC#BM 241102309)". PATIENT OUTCOME: "IT IS UNKNOWN WHETHER THERE WAS ANY HEALTH HARM TO THE PATIENT'S HEALTH."
Patient problems: No Clinical Signs, Symptoms or Conditions; Thromboembolism
MAUDE report 20943883 ↗
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record