TREO Abdominal Stent-Graft System 28-B2-26-120URecalls & reported problems
Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant
All models of TREO Abdominal Stent-Graft System (73)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00843576152519), received between Jan 3, 2025 and Jul 8, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Separation” (1 report).
1 other recall cover devices cleared under the same FDA submission (P190015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult or Delayed Separation1
- Output Problem1
- Positioning Problem1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2025Output Problem; Positioning Problem
GRAFT WAS OPENED AND MODIFIED. DIFFICULTY RESHEATHING GRAFT.
Patient problems: Insufficient Information
MAUDE report 22436178 ↗MalfunctionJan 3, 2025Difficult or Delayed Separation; Difficult or Delayed Separation
"DESCRIBE THE EVENT OR PROBLEM: DURING THE CASE, THE TREO ABDOMINAL STENT-GRAFT SYSTEM WAS STUCK ON DELIVERY. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? FENESTRATED ENDOVASCULAR REPAIR OF VISCERAL AND INFRARENAL AORTA PERCUTANEOUS INCLUDING FOUR OR MORE VISCERAL ARTERY ENDOPROSTHESES WITH BIFURCATE. WHAT PROBLEM DID THE USER HAVE: DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING). ADDITIONAL INFORMATION RECEIVED ON 12/26/2024: THE DEVICE WAS IMPLANTED AND THE PATIENT IS FINE. THE DEVICE IS NOT AVAILABLE FOR RETURN. THE PROXIMAL CLASP MOVED DOWN BUT DID NOT RELEASE THE SUPRARENAL FIXATION. THE IMAGING I HAVE AVAILABLE ALONG WITH DR. (B)(6) CASE PLAN. I WILL HAVE TO SEND THOSE IN BATCHES; THE FILE IS TOO LARGE. THIS WAS A PHYSICIAN MODIFIED DEVICE." PATIENT OUTCOME: "THE PATIENT IS FINE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21072079 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P190015. One submission often covers many models, so check the product and lot details in each record.
- Z-1689-2023 · May 3, 2023 · Open, ClassifiedThe TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graf
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page