Skip to content

TREO Abdominal Stent-Graft System 28-B2-28-120U (UDI 00843576152526)Recalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of TREO Abdominal Stent-Graft System (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576152526), received between Nov 8, 2024 and Nov 8, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Device Endoleak” (1 report).

1 other recall cover devices cleared under the same FDA submission (P190015); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Endoleak1
  • Failure to Seal1

Patient problems

  • Clinical Endoleak1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 8, 2024Adverse Event Without Identified Device or Use Problem; Failure to Seal; Device Endoleak; Adverse Event Without Identified Device or Use Problem; Failure to Seal; Device Endoleak

    "DOCTOR NOTIFIED ME THAT THE FOLLOW UP SCAN HAS AN ENDOLEAK. ADDITIONAL INFORMATION RECEIVED ON 10/17/2024: PROCEDURE WAS DONE YESTERDAY. WE NEVER DID AN ANGIO TO TRULY SEE IF THERE WAS A TYPE 1A VS WHAT I BELIEVE IS A TYPE 2. THE DR. ADMITTED THERE WAS A PROBABLY A TYPE 2. BUT WANTED TO TREAT AS THOUGH THERE WAS A POSSIBLE 1A ALSO. HE ENDED UP PUTTING IN 6 ENDO ANCHORS AND THEN ADDED A 28X40 TREO CUFF. PATIENT IS DOING WELL LAST I HEARD." PATIENT OUTCOME: "PATIENT SCHEDULED FOR FOLLOW UP PROCEDURE TO DETERMINE SOURCE OF ENDOLEAK AND TREAT ON (B)(6) 2024"

    Patient problems: Clinical Endoleak

    MAUDE report 20640423 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P190015. One submission often covers many models, so check the product and lot details in each record.

  • Z-1689-2023 · May 3, 2023 · Open, Classified
    The TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graf

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page