Skip to content

TREO Abdominal Stent-Graft System 28-C2-30-055URecalls & reported problems

Bolton Medical Inc. · Endovascular Stent-Graft System · Class 3 · Implant

All models of TREO Abdominal Stent-Graft System (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00843576152694), received between Nov 25, 2024 and Nov 25, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Device Endoleak” (1 report).

1 other recall cover devices cleared under the same FDA submission (P190015); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Endoleak1
  • Failure to Seal1

Patient problems

  • Clinical Endoleak1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2024Failure to Seal; Device Endoleak; Failure to Seal; Device Endoleak

    "EVALUATED A CT SCAN FOR DR. (B)(6) ON (B)(6) 2024. ENDOLEAK WAS NOTED ON THE SCAN. POTENTIAL TYPE 3, 1A OR 2. SOURCE WAS TO BE DETERMINED POST SELECTIVE ANGIOGRAMS SCHEDULED ON (B)(6) 2023. PER MY DISCUSSION WITH DR. (B)(6), THE INITIAL TREO GRAFT WAS IMPLANTED 1CM DISTAL TO THE TAKEOFF THE LRA IN 2021. DURING THE INITIAL ANGIOGRAM PERFORMED ON (B)(6) 2025, BRISK FILLING ON THE ANEURYSM SAC WAS NOTED. DR. (B)(6) MADE THE DECISION TO PLACE A TREO CUFF TO THE LEVEL OF THE LRA (30MM X 55MM). THEREAFTER, IT WAS MOLDED USING THE CODA BALLOON AND ANOTHER ANGIOGRAM WAS TAKEN. THIS ANGIO STILL REVEALED THE LEAK. THEN THE DECISION WAS MADE TO LEAVE THE CODA BALLOON INFLATED IN THE PROXIMAL SEGMENT OF THE GRAFT/CUFF AND PLACE THE PIGTAIL CATHETER DISTAL TO THE BALLOON AND ABOVE THE FLOW-DIVIDER OF THE GRAFT. DURING THIS INJECTION, WE NOTED THE JETTING OF CONTRAST ON THE RIGHT MEDIAL SIDE OF THE TREO MAIN BODY GRAFT THROUGH THE FABRIC. THIS CONFIRMED THE TYPE 3B ENDOLEAK. DR. (B)(6) DECIDED TO CONFIRM THIS FINDING BY TAKING A BERNSTEIN CATHETER AND WAS ABLE TRAVERSE THE CATHETER THROUGH THE FABRIC TEAR OF THE GRAFT AND THEREAFTER PLACING A WIRE THROUGH THE CATHETER INTO THE ANEURYSM SAC. WAI

    Patient problems: Clinical Endoleak

    MAUDE report 20771515 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P190015. One submission often covers many models, so check the product and lot details in each record.

  • Z-1689-2023 · May 3, 2023 · Open, Classified
    The TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graf

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page