MasimoSET LNCS® TC-I SpO2 Reusable Ear Sensor 1895Recalls & reported problems
Masimo Corporation · MasimoSET LNCS TC-I SpO2 Reusable Ear Sensor · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843997000048), received between Feb 5, 2019 and Nov 13, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Readings” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K051212); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 21, 2023Class IIOpen, Classified
TC-I tip clip sensor may produce readings outside the accuracy specification.
Z-1985-2023 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- High Readings1
- Incorrect Measurement1
- Low Readings1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 13, 2025High Readings; High Readings
THE CUSTOMER REPORTED "A PATIENT HOSPITALIZED FOR TYPE 2 RESPIRATORY FAILURE AND VENTILATED WITH NIV PRESENTED A MARKED DISCREPANCY BETWEEN THE MEASURED SATURATION VALUES(SPO2 89¿91%) AND THE RESULTS OF SEVERAL SUCCESSIVE BLOOD GAS ANALYSES, INDICATING SEVERE HYPOXEMIA. CLINICAL CHECKS SHOWED NO DETERIORATION, AND THE SPO2 SIGNAL APPEARED TO BE OF ADEQUATE QUALITY (PERFUSION INDEX 1.25). PER THE REPORTED INFORMATION, THE PATIENT DID NOT EXPERIENCE ANY COMPLICATIONS NOR SERIOUS CONSEQUENCES.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 23550494 ↗MalfunctionMar 4, 2023Incorrect Measurement; Incorrect Measurement
THE CUSTOMER REPORTED THE SENSOR "SHOWS VALUES THAT SEEM CORRECT, AND THE NEXT TIME YOU GET DIFFERENT VALUES." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16483290 ↗MalfunctionFeb 5, 2019Low Readings; Low Readings
THE CUSTOMER REPORTED THE DEVICE PROVIDED INACCURATE SPO2 VALUES. THE EXPECTED VALUE WAS 94 AND THE MONITOR SHOWED 74. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8308621 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051212. One submission often covers many models, so check the product and lot details in each record.
- Z-1986-2023 · Apr 21, 2023 · Open, ClassifiedMasimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 2503
- Z-2640-2017 · Apr 4, 2016 · TerminatedSpO2 connecting cable for Masimo LNCS sensor (Merge Part Number: HW-HEMO-00014), Product Usage: Merge Hemodynamics monitors, measures, and records physiologic data from a human pa
- Z-2642-2017 · Apr 4, 2016 · TerminatedMasimoSET LNOP DCIP pediatric/slender digit SpO2 reusable sensor, finger clip pediatric (Merge Part Number: HW-HEMO-00076), Product Usage: Merge Hemodynamics monitors, measures,
- Z-2641-2017 · Apr 4, 2016 · TerminatedMasimo SPO2 Sensor, ear clip-adult (Merge Part Number: HW-HEMO-00047), Product Usage: Merge Hemodynamics monitors, measures, and records physiologic data from a human patient und
Other Oximeter, ear devices
- MasimoSET® M-LNCS™ TC-I SpO2 Reusable Ear Sensor 2503Masimo Corporation · Oximeter, ear1 recall1 report
- Portrait VSM MVC101Ge Medical Systems Information Technologies, Inc. · Oximeter, ear1 recall0 reports
- TruSignal SpO2 TS-SE-3GE Healthcare Finland Oy · Oximeter, ear1 recall0 reports
- V-Sign™ Sensor 2 VS-A/P/NSenTec AG · Oximeter, Ear1 recall0 reports
- MasimoSET® LNOP® TC-I SpO2 Reusable Ear Sensor 1794Masimo Corporation · Oximeter, ear1 recall0 reports
- Membrane Changer (single-use) / single unit MCSenTec AG · Oximeter, Ear1 recall0 reports
- TruSignal SpO2 TS-APGE Healthcare Finland Oy · Oximeter, ear1 recall0 reports
- TruSignal SpO2 TS-W-DGE Healthcare Finland Oy · Oximeter, ear1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page