Skip to content

19g Packer/Chang IOL Cutters DFH-0012Recalls & reported problems

Microsurgical Technology, Inc. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00844685001323), received between Oct 8, 2018 and Apr 12, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1
  • Device Handling Problem1
  • Material Fragmentation1
  • Mechanical Problem1
  • Naturally Worn1

Patient problems

  • Capsular Bag Tear1
  • Cataract1
  • No Clinical Signs, Symptoms or Conditions1
  • Tissue Damage1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 12, 2024Naturally Worn; Naturally Worn

    SCISSOR TIPS DON'T FULLY CLOSE, SO THE SURGEON HAD TO CUT THE IOL MORE TOWARDS THE BASE OF THE SCISSORS. BECAUSE OF THIS, THE TIPS WERE FURTHER EXTENDED WITHIN THE ANTERIOR CHAMBER, AND THEY MADE CONTACT WITH/RUPTURED THE CAPSULAR BAG, WHICH PROMPTED THE NEED FOR AN ADDITIONAL PROCEDURE.

    Patient problems: Capsular Bag Tear

    MAUDE report 19102234 ↗
  • MalfunctionAug 3, 2023Degraded; Material Fragmentation; Degraded; Material Fragmentation

    WE HAD THE SCISSORS BREAK IN THE EYE TODAY WHILE CUTTING AN ACRYLIC LENSE. THE BROKEN PARTS WERE SUCCESSFULLY REMOVED FROM THE EYE. PATIENT FOLLOW UP THE NEXT DAY PER USUAL RETINA PROCEDURE. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17448168 ↗
  • InjuryOct 8, 2018Mechanical Problem; Device Handling Problem; Mechanical Problem; Device Handling Problem

    A SCISSOR WAS BEING USED TO CUT AN IOL WHEN A HOLE IN THE ANTERIOR CHAMBER OCCURRED.

    Patient problems: Cataract; Tissue Damage

    MAUDE report 7943472 ↗

Other Scissors, Ophthalmic devices

All Scissors, Ophthalmic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page