ALPHATEC SOLUS 25909Recalls & reported problems

Alphatec Spine, Inc. · ALPHATEC SOLUS DISTAL WRENCH TIP · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00844856062719), received between Jul 31, 2018 and Jul 31, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2018Break

    THE DISTAL TIP OF THE INSTRUMENT BROKE OFF IN THE IMPLANT WHEN THE SURGEON WAS TURNING THE DEPLOYMENT WRENCH HANDLES. THE TIP REMAINS WITHIN THE IMPLANT POSITIONED BETWEEN THE L5-S1. ALTHOUGH THE PATIENT HAD VERY HARD SCLEROTIC BONE, BOTH BLADES OF THE IMPLANT WERE ABLE TO BE COMPLETELY DEPLOYED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7735893 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices