Trestle Luxe 71002-034Recalls & reported problems
Alphatec Spine, Inc. · ANTERIOR CERVICAL PLATE ASSEMBLY, 2-LEVEL, 034mm · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00844856067011), received between Mar 1, 2017 and Nov 18, 2020 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K102820); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated3
- Material Twisted/Bent2
- Break1
- Component Falling1
- Detachment Of Device Component1
- Detachment of Device or Device Component1
- Material Protrusion/Extrusion1
- Use of Device Problem1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient2
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2020Break
UPON USING THE PLATE BENDING, THE LOCKING MECHANISM BROKE OFF THE PLATE.
Patient problems: Insufficient Information
MAUDE report 10861201 ↗InjuryAug 10, 2020Detachment of Device or Device Component; Material Protrusion/Extrusion
REVISION SURGERY WAS CONDUCTED ON (B)(6) 2020 TO REMOVE AND REPLACE A 2-LEVEL TRESTLE LUXE PLATE. THE LOCKING MECHANISM BECAME DETACHED WHICH ALLOWED THE SCREW TO PROTRUDE. THE TRESTLE LUXE ANTERIOR CERVICAL PLATING SYSTEM WAS INITIALLY IMPLANTED ON (B)(6) 2020.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10385413 ↗MalfunctionApr 24, 2019Device Dislodged or Dislocated; Material Twisted/Bent
THE GOLD PIECE IN THE CENTER OF THE PLATE POPPED OFF AFTER BENDING TO ACHIEVE ADDITIONAL LORDOSIS. NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8544614 ↗MalfunctionApr 24, 2019Device Dislodged or Dislocated; Material Twisted/Bent
THE GOLD PIECE IN THE CENTER OF BOTH PLATES POPPED OFF AFTER BENDING EACH OF THEM TO ACHIEVE ADDITIONAL LORDOSIS. AN ADDITIONAL PLATE WAS ALSO BENT TO FIT THE PATIENT. THE GOLD PIECE POPPED OFF THIS ONE AS WELL, BUT THE SURGEON ELECTED TO INSTALL THE PLATE WITHOUT A LOCKING MECHANISM. THE EVENT CAUSED NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8544492 ↗MalfunctionApr 24, 2018Device Dislodged or Dislocated
THE SURGEON GRASPED A PLATE WITH PLATE HOLDER AND AS HE WAS TRYING IT IN PATIENT, ONE OF THE LOCKING SLIDES POPPED OFF. SURGEON INSTALLED THE PLATE WITHOUT A LOCKING MECHANISM AT THAT POSITION, LEAVING THE HOLE EMPTY (C6).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7454600 ↗MalfunctionMar 1, 2017Detachment Of Device Component; Component Falling; Use of Device Problem
DURING THE CASE, A TRESTLE LUXE 2-LEVEL PLATE WAS RENDERED USELESS AS THE MIDDLE "LOCKING BAR" FELL OUT AS THE SURGEON ATTEMPTED TO BEND THE PLATE. THERE WERE NO PATIENT COMPLICATIONS. A SECOND 34MM PLATE WAS SELECTED, BENT APPROPRIATELY AND IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6370041 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K102820. One submission often covers many models, so check the product and lot details in each record.
- Z-1547-2015 · Apr 10, 2015 · Terminated4.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm Alphatec Spine Solutions for the Aging Spine Trestle Luxe Anterior Cervical Plating System REF 71240-14 LOT XXXXXX Alphat
- Z-2273-2012 · May 29, 2012 · TerminatedTRESTLE LUXE Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cerv
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page