Trestle Luxe 71004-064Recalls & reported problems
Alphatec Spine, Inc. · ANTERIOR CERVICAL PLATE ASSEMBLY, 4-LEVEL, 064mm · Class 2 · Implant
All models of Trestle Luxe (24)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00844856067189), received between Mar 13, 2017 and Feb 5, 2020 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K102820); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Device Operates Differently Than Expected1
- Material Protrusion/Extrusion1
- Material Twisted/Bent1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2020Detachment of Device or Device Component; Material Twisted/Bent
TWO 64MM PLATES WERE BENT DURING SURGERY AND GOLD PIECES SNAPPED OFF. THE SURGEON INDICATED THAT WHEN HE BENT THE PLATES, THE GOLD SLIDES DID NOT COME OFF UNTIL INSERTING THE SCREWS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9672072 ↗InjuryNov 30, 2018Material Protrusion/Extrusion
DURING A REVISION CASE, THE SURGEON PLACED ALL SCREWS INTO PLATE. ONE OF THE SCREWS WENT THROUGH THE PLATE CAUSING REMOVAL OF THE ENTIRE CONSTRUCT. THE EVENT CAUSED THE SURGERY TO BE DELAYED APPROXIMATELY 1.5 TO 2 HOURS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8120883 ↗InjuryMar 13, 2017Device Operates Differently Than Expected
WHILE INSERTING A 4-LEVEL TRESTLE LUXE PLATE ONE OF THE SELF-TAPPING SCREWS PASSED COMPLETELY THRU THE PLATE. THE PLATE WAS REMOVED AND REPLACED WITH A NEW 3-LEVEL PLATE AT WHICH TIME THE RESCUE SCREW ALSO PASSED COMPLETELY THRU THIS NEWLY IMPLANTED PLATE. THE RESCUE SCREW WAS REMOVED FROM THE PATIENT WHILE THE 3-LEVEL TRESTLE LUXE PLATE REMAINS SECURED WITHIN THE PATIENT'S CERVICAL SPINE. REMOVAL AND REPLACEMENT OF THE IMPLANTS CAUSED A 30-40 MINUTE DELAY IN THE CASE.
Patient problems: No Code Available
MAUDE report 6401290 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K102820. One submission often covers many models, so check the product and lot details in each record.
- Z-1547-2015 · Apr 10, 2015 · Terminated4.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm Alphatec Spine Solutions for the Aging Spine Trestle Luxe Anterior Cervical Plating System REF 71240-14 LOT XXXXXX Alphat
- Z-2273-2012 · May 29, 2012 · TerminatedTRESTLE LUXE Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cerv
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page