Skip to content

OSSEOTITE® CERTAIN® IOSS485Recalls & reported problems

ZimVie US Corp LLC · OSSEOTITE® CERTAIN® IMPLANT 4 X 8.5MM · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
21
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 00844868007142), received between Aug 12, 2020 and Aug 16, 2026 — classified as 21 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of Osseointegration” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 4 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem20
  • Loss of Osseointegration4
  • Fracture1
  • Mechanical Problem1

Patient problems

  • Unspecified Infection20
  • Osteolysis18
  • No Code Available2
  • Abscess1
  • Inflammation1
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1
  • Sinus Perforation1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 16, 2026Adverse Event Without Identified Device or Use Problem

    Doctor reported that implant was removed due to Peri-implantitis with complete loss of integration.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 50268120 ↗
  • InjuryMay 19, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #26 WAS REMOVED DUE TO PERI-IMPLANTITIS WITH TOTAL LOSS OF INTEGRATION.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 25221614 ↗
  • InjuryApr 21, 2026Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANTS AT TOOTH SITES 16, 15 WERE REMOVED DUE TO PERI-IMPLANTITIS WITH TOTAL LOSS OF INTEGRATION. PATIENT HAS BEEN SCHEDULED FOR NEW IMPLANT PLACEMENT IN APROXIMATELY 3 MONTHS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24936785 ↗
  • InjuryApr 21, 2026Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANTS AT TOOTH SITES 16, 15 WERE REMOVED DUE TO PERI-IMPLANTITIS WITH TOTAL LOSS OF INTEGRATION. PATIENT HAS BEEN SCHEDULED FOR NEW IMPLANT PLACEMENT IN APPROXIMATELY 3 MONTHS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24936682 ↗
  • InjuryApr 18, 2026Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 26 WAS REMOVED DUE TO PERI-IMPLANTITIS WITH TOTAL LOSS OF INTEGRATION. PATIENT HAS BEEN SCHEDULED FOR NEW IMPLANT PLACEMENT IN APPROXIMATELY 3 MONTHS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24919744 ↗
  • InjuryApr 18, 2026Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 45 WAS REMOVED DUE TO PERI-IMPLANTITIS WITH TOTAL LOSS OF INTEGRATION. PATIENT HAS BEEN SCHEDULED FOR NEW IMPLANT PLACEMENT IN APPROXIMATELY 3 MONTHS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24919743 ↗
  • InjuryMar 5, 2026Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED SCREW FRACTURE AT TOOTH SITE 26 AND PERI-IMPLANTITIS WITH TOTAL LOSS OF INTEGRATION. PATIENT HAS BEEN RESCHEDULED FOR NEW IMPLANT PLACEMENT IN APPROX. 3 MONTHS. THIS REPORT REFERS TO PERI-IMPLANTITIS EVENT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24517496 ↗
  • InjuryMar 3, 2026Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24500200 ↗
  • InjuryApr 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 17 WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 21888670 ↗
  • InjuryJul 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 11 WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 19663487 ↗
  • InjuryJun 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 16 WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 19523755 ↗
  • InjuryMay 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 26 WAS REMOVED DUE TO PERI-IMPLANTITIS. PATIENT HAS BEEN RESCHEDULED FOR BONE REGENERATION.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 19403990 ↗
  • InjuryMar 20, 2024Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem; Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Osteolysis; Sinus Perforation; Unspecified Infection

    MAUDE report 18942819 ↗
  • MalfunctionDec 15, 2023Mechanical Problem; Mechanical Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18338483 ↗
  • InjuryOct 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED PERI-IMPLANTITIS AT TOOTH SITE 26. A NEW IMPLANT WILL BE PLACED IN 3 MONTHS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 17910654 ↗
  • InjuryOct 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PERI-IMPLANTITIS. PATIENT HAS BEEN RESCHEDULED FOR NEW IMPLANT PLACEMENT IN 3 MONTHS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 15536123 ↗
  • InjurySep 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED PERI-IMPLANTITIS AT TOOTH SITE 26. PATIENT HAS BEEN RESCHEDULED FOR NEW IMPLANT PLACEMENT IN APROX. 3 MONTHS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 15463833 ↗
  • InjuryAug 30, 2022Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem; Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO PERI-IMPLANTITIS AT THE SITE WITH A LOSS OF INTEGRATION.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 15320566 ↗
  • InjuryApr 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED PERI-IMPLANTITIS WITH INFLAMMATION.

    Patient problems: Inflammation; Osteolysis; Unspecified Infection

    MAUDE report 14236610 ↗
  • InjuryFeb 12, 2021Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Information

    MAUDE report 11320834 ↗
  • InjurySep 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED IMPLANT REMOVED DUE TO PERI-IMPLANTITIS. PATIENT ALSO HAD ABCESS.

    Patient problems: Abscess; No Code Available; Unspecified Infection

    MAUDE report 10498689 ↗
  • InjuryAug 12, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED IMPLANT REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: No Code Available; Unspecified Infection

    MAUDE report 10395192 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page