OSSEOTITE® CERTAIN® XIFNT410Recalls & reported problems
ZimVie US Corp LLC · OSSEOTITE® TAPERED CERTAIN® IMPLANT 4 X 10MM · Class 2 · Implant
All models of OSSEOTITE CERTAIN (74)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This ZimVie Osseotite Certain implant has 194 reports, 186 injuries and no deaths. Many reports describe infection, inflammation and bone loss around the implant, and several say it had to be removed. Some describe pain or fracture. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 193
- Deaths reported
- 0
- Injuries reported
- 185
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 193 adverse event reports that list this exact device (DI 00844868021957), received between Oct 18, 2019 and Sep 30, 2026 — classified as 185 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Fracture” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem170
- Fracture18
- Mechanical Problem5
- Loss of Osseointegration3
- Positioning Failure1
- Unsealed Device Packaging1
Patient problems
- Unspecified Infection134
- Osteolysis96
- Pain46
- Inflammation36
- No Clinical Signs, Symptoms or Conditions25
- Swelling/ Edema11
- No Information10
- No Code Available8
Recent adverse event reports
The 25 most recent of 193 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem
It was reported loss of integration due to Peri-implantitis at tooth site 21.
Patient problems: Osteolysis; Unspecified Infection
MAUDE report 50944161 ↗InjuryJul 28, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the dental implant at Tooth Location 21 failed due to peri-implantitis. Pain and inflammation were also reported.
Patient problems: Inflammation; Osteolysis; Unspecified Infection
MAUDE report 50051765 ↗InjuryJun 11, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #8 WAS PLACED WITH HYBRID PLANNED. THE IMPLANT DEVELOPED PERI-IMPLANTITIS AND HAD TO BE REMOVED.
Patient problems: Osteolysis; Unspecified Infection
MAUDE report 25472953 ↗InjuryJun 9, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #36 WAS REMOVED DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 25448212 ↗InjuryJun 9, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #16 WAS REMOVED DUE TO INFECTION. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.
Patient problems: Inflammation; Unspecified Infection
MAUDE report 25448015 ↗InjuryJun 9, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #36 WAS REMOVED DUE TO PERI-IMPLANTITIS.
Patient problems: Osteolysis; Unspecified Infection
MAUDE report 25447209 ↗InjuryMay 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THE IMPLANTS AT TOOTH SITE #45-46 WERE REMOVED DUE TO PERIIMPLANTITIS . PATIENT REFERRED PAIN. NEW IMPLANTS WILL BE PLACED AFTER REGENERATION.
Patient problems: Osteolysis; Pain; Unspecified Infection
MAUDE report 25122833 ↗InjuryApr 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #13 WAS REMOVED DUE TO INFECTION. IMMEDIATE IMPLANT WAS PLACED WITH ALLOGRAFT AT 20 N CM AND COVER SCREW WAS PLACED. AT THREE WEEKS THERE WAS A SUTURE REMOVAL APPOINTMENT. THE IMPLANT IMPLANT WAS LOOSE AND WAS REMOVED WITH TWEEZERS. THE IMPLANT SITE WAS DEGRANULATED AND A RIDGE PRESERVATION WAS COMPLETED WITH PTFE MEMBRANE. SYMPTOMS AS A RESULT OF THE EVENT: EDEMA AND INFLAMMATION.
Patient problems: Inflammation; Swelling/ Edema; Unspecified Infection
MAUDE report 25000501 ↗InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AN IMPLANT AT TOOTH SITE #13 WAS REMOVED DUE TO A PERI-IMPLANTITIS PATIENT SUFFERED FROM PAIN AND INFLAMMATION.
Patient problems: Inflammation; Osteolysis; Unspecified Infection
MAUDE report 24963945 ↗InjuryMar 10, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #25 WAS REMOVED DUE TO BONE LOSS AFTER SINUS LIFT. THE PROCEDURE WAS NOT COMPLETED WITH ANOTHER IMPLANT.
Patient problems: Osteolysis
MAUDE report 24560272 ↗InjuryMar 6, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #21 WAS REMOVED DUE TO PERI-IMPLANTITIS.
Patient problems: Osteolysis; Unspecified Infection
MAUDE report 24536528 ↗MalfunctionFeb 13, 2026Mechanical Problem
DOCTOR REPORTS WHEN TRYING TO PLACE THE IMPLANT AT TOOTH LOCATION 46, THE DRIVER SPINS FREELY AND DISENGAGES FROM THE IMPLANT. ANOTHER IMPLANT WAS PLACED DURING THE SAME PROCEDURE USING THE SAME DRIVER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24330373 ↗InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.
Patient problems: Osteolysis; Pain; Unspecified Infection
MAUDE report 24328269 ↗InjuryFeb 10, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #36 WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS AND SUPPURATION.
Patient problems: Erosion; Osteolysis; Unspecified Infection
MAUDE report 24306075 ↗InjuryJan 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT IMPLANT WAS REMOVED DUE TO PERI IMPLANTITIS. PATIENT PRESENTED FOR INTEGRATION CHECK IMPLANT SITE 10. BONE LOSS ON PANO. PATIENT COMPLAINTS OF PAIN. TISSUE IS RED. IMPLANT REMOVED. SITE WAS BONE GRAFTED AND PATIENT WILL RETURN IN 4 MONTHS FOR REPLACEMENT. NOTED PAIN AND INFLAMMATION. WILL RETURN IN 4 MONTHS FROM REPLACEMENT.
Patient problems: Osteolysis; Pain; Unspecified Infection
MAUDE report 24193468 ↗InjuryJan 16, 2026Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24091503 ↗InjuryJan 15, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #36 WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.
Patient problems: Inflammation; Osteolysis; Unspecified Infection
MAUDE report 24085190 ↗InjuryJan 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #26 WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.
Patient problems: Inflammation; Osteolysis; Unspecified Infection
MAUDE report 24015328 ↗InjuryDec 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #23 WAS REMOVED DUE TO BONE LOSS. PATIENT EXPERIENCED INFLAMMATION.
Patient problems: Inflammation; Osteolysis
MAUDE report 23858503 ↗InjuryDec 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 45 WAS REMOVED DUE TO PERI-IMPLANTITIS. AFTER 3-4 WEEKS OF CROWN PLACEMENT, PATIENT CAME TO CLINICIAN INDICATING CROWN MOBILITY, BUT DOCTOR NOTICED THAT IT WAS THE IMPLANT INSTEAD. PATIENT REFERRED PAIN.
Patient problems: Osteolysis; Pain; Unspecified Infection
MAUDE report 23756878 ↗InjuryDec 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO BONE LOSS. NO OTHER IMPLANT WAS PLACED DURING THE SAME PROCEDURE.
Patient problems: Osteolysis
MAUDE report 23733325 ↗InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTS WERE REMOVED DUE TO PERI-IMPLANTITIS. THE PROCEDURE WAS NOT COMPLETED WITH OTHER IMPLANTS.
Patient problems: Osteolysis; Unspecified Infection
MAUDE report 23720501 ↗InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTS WERE REMOVED DUE TO PERI-IMPLANTITIS. THE PROCEDURE WAS NOT COMPLETED WITH OTHER IMPLANTS.
Patient problems: Osteolysis; Unspecified Infection
MAUDE report 23720381 ↗MalfunctionNov 10, 2025Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23510696 ↗InjuryNov 3, 2025Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #16 FRACTURED AT THE COLLAR. IMPLANT WAS REMOVED. BONE LOSS WAS ALSO REPORTED.
Patient problems: Osteolysis
MAUDE report 23448432 ↗
FDA clearance and classification
The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the OSSEOTITE® CERTAIN® XIFNT410
Has the OSSEOTITE® CERTAIN® XIFNT410 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00844868021957), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the OSSEOTITE® CERTAIN® XIFNT410?
The FDA's MAUDE database lists 193 adverse event reports for this device, received between Oct 18, 2019 and Sep 30, 2026. In those reports, the most common specific device problems are “Fracture” (18), “Mechanical Problem” (5) and “Loss of Osseointegration” (3); the most common patient problems are “Unspecified Infection” (134), “Osteolysis” (96) and “Pain” (46). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the OSSEOTITE® CERTAIN® XIFNT410?
None of the reports are classified as deaths, 185 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the OSSEOTITE® CERTAIN® XIFNT410 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the OSSEOTITE® CERTAIN® XIFNT410, and is it still sold?
The FDA's device database lists ZimVie US Corp LLC as the company responsible for it (brand OSSEOTITE® CERTAIN®, model XIFNT410). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,942 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,651 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,576 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,756 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,548 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,631 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,501 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,423 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page