OSSEOTITE® CERTAIN® PREVAIL XIITP6511Recalls & reported problems

ZimVie US Corp LLC · OSSEOTITE® TAPERED CERTAIN® PREVAIL® IMPLANT 6/5 X 11.5 MM · Class 2 · Implant

All models of OSSEOTITE CERTAIN PREVAIL (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This is a ZimVie OSSEOTITE dental implant, a small screw-shaped device placed in the jaw to hold a tooth replacement. The 50 reports mostly describe implants that became loose or were removed after infection around the implant, with bone loss and pain;45 reports involved injuries and no deaths were reported. A few reports mention mechanical problems with the implant itself. The number of reports held steady at 6 over each of the past two 12-month periods, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
45
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 00844868022213), received between Dec 2, 2019 and Apr 1, 2026 — classified as 45 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem43
  • Mechanical Problem7

Patient problems

  • Unspecified Infection22
  • Osteolysis14
  • No Clinical Signs, Symptoms or Conditions12
  • Pain12
  • Abscess3
  • No Code Available3
  • No Information3
  • Swelling/ Edema2

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 1, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY MAIL THAT THE IMPLANT IN SITE 3 WAS MOBILE AND FAILED DUE TO PERI-IMPLANTITIS. PATIENT WILL RETURN FOR NEW IMPLANT AND BONE GRAFT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24758383 ↗
  • InjuryMar 16, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #3 WAS MOBILE. THE IMPLANT HAD FAILED AND WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24605033 ↗
  • InjuryFeb 27, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #3 WAS MOBILE AND FAILED DUE TO PERI-IMPLANTITIS. THE IMPLANT WAS REMOVED.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24470617 ↗
  • InjuryDec 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS BONE LOSS WITH TISSUE RECESSION AROUND THE IMPLANT. THE IMPLANT WAS REMOVED. TOOTH LOCATION IS UNKNOWN.

    Patient problems: Osteolysis

    MAUDE report 23922158 ↗
  • MalfunctionDec 9, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT PATIENT HAD CROWN PLACED AND WAS MOBILE, CROWN REMOVED AND IMPLANT STABLE BUT PAIN. CROWN REPLACED BUT STILL MOBILE IMPLANT REMOVED AND REPLACED. TOOTH SITE: 30. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Pain

    MAUDE report 23757866 ↗
  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED CONSTANT PAIN AT THE IMPLANT AT TOOTH SITE #19 AND CONTINUED TO HAVE PAIN AND A COLD SENSATION IN THEIR LOWER JAW. AN ADDITIONAL APPOINTMENT WAS REQUIRED TO REPLACE THE IMPLANT. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Pain

    MAUDE report 23692786 ↗
  • InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23311228 ↗
  • InjurySep 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #14 WAS REMOVED DUE TO INFECTION. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 23071742 ↗
  • InjuryApr 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT IMPLANT NEEDED TO BE PLACED DEEPER TO ALLOW SPACE FOR IMPLANT CROWN, PROPER OCCLUSION. TOOTH NUMBER 31. IMPLANT REPLACED WITH A DIFFERENT IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21909119 ↗
  • InjuryFeb 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #30 WAS REMOVED DUE TO BONE LOSS.

    Patient problems: Osteolysis

    MAUDE report 21377181 ↗
  • InjuryDec 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 WAS REMOVED DUE TO BONE LOSS. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.

    Patient problems: Inflammation; Osteolysis

    MAUDE report 20932600 ↗
  • InjuryNov 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT IMPLANT WAS REMOVED DUE TO INFECTION. GRANULATION TISSUE AROUND IMPLANT CAUSING PAIN AND NON HEALING. PER INDICATED: SYMPTOMS AS A RESULT OF THE EVENT: PAIN, INFLAMMATION. SURGICAL/MEDICAL INTERVENTION REQUIRED FOR PERMANENT DAMAGE: YES, REMOVAL. WAS THERE A DELAY DURING THE PROCEDURE: NO. ADDITIONAL APPOINTMENT REQUIRED: NO. WAS THE PROCEDURE COMPLETED USING ANOTHER IMPLANT OR ANOTHER DEVICE: NO.

    Patient problems: Unspecified Infection

    MAUDE report 20650969 ↗
  • InjurySep 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 20257106 ↗
  • InjuryAug 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Unspecified Infection

    MAUDE report 19959246 ↗
  • InjuryJul 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #14 WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 19700825 ↗
  • InjuryApr 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL OR CORRECTED INFORMATION TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19160028 ↗
  • InjuryApr 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DOCTOR TRIED TO PLACE THE IMPLANT AT TOOTH LOCATION #18, HOWEVER A POSSIBLE BUCCAL FRACTURE WAS NOTED BECAUSE THE TORQUE WAS GOOD THEN IT WASN'T. THE IMPLANT WAS UNABLE TO BE PLACED AND THE SOCKET WAS GRAFTED. THE PATIENT WILL RETURN TO THE CLINIC IN FOUR (4) MONTHS TO PLACE A NEW IMPLANT.

    Patient problems: Bone Fracture(s)

    MAUDE report 19143179 ↗
  • InjuryMar 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 LOST INTEGRATION DUE TO PERI-IMPLANTITIS AND WAS REMOVED.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 18857074 ↗
  • InjuryDec 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PERI-IMPLANTITIS, TOOTH14. SYMPTOMS AS A RESULT OF THE EVENT: BONE LOSS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 18375512 ↗
  • MalfunctionDec 4, 2023Mechanical Problem; Mechanical Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18256712 ↗
  • InjuryAug 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #18 LOST INTEGRATION DUE TO PERIIMPLANTITIS AND WAS REMOVED AND BONE GRAFTED. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 17646946 ↗
  • MalfunctionAug 28, 2023Mechanical Problem; Mechanical Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17631624 ↗
  • InjuryJun 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT TOOTH SITE # 19 WAS INFECTED AND THE PATIENT HAD PAIN. THE IMPLANT WAS REMOVED, AND BONE GRAFT PLACED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 17080683 ↗
  • InjuryMay 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER REPORTED REMOVAL OF IMPLANT DUE TO IMPLANT POSITIONING. UNABLE TO RESTORE WITH IMPLANT CROWN. REMOVED AND REPLACEMENT. TOOTH 18 (UNIVERSAL).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16883746 ↗
  • InjuryMay 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED PERI-IMPLANTITIS AND LOSS OF INTEGRATION AROUND THE IMPLANT AT TOOTH LOCATION #30. THE IMPLANT WAS REMOVED. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 16883032 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices

All IMPLANT, ENDOSSEOUS, ROOT-FORM devices →