TSX Implants TSX37B13Recalls & reported problems

ZimVie US Corp LLC · TSX™ IMPLANT, 3.7MMD, 13MML · Class 2 · Implant

Quick summary

This is a ZimVie TSX dental implant, a small screw-shaped root implant used to replace missing teeth. Most of the 75 reports describe implants that loosen and fail to stay in the jaw, with infection, bone loss around the implant, pain and inflammation around the gum; several mention implants having to be removed and bone grafts done. The reports describe 64 injuries and no deaths. Some also mention packaging that was not sealed and parts missing from the package. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
0
Injuries reported
59
Recalls naming it
0

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 00844868045069), received between Jun 29, 2023 and Sep 15, 2026 — classified as 59 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Unsealed Device Packaging” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 4 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 5 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 7 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 5 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem59
  • Unsealed Device Packaging5
  • Component Missing3
  • Separation Failure2
  • Mechanical Problem1

Patient problems

  • Unspecified Infection46
  • Osteolysis36
  • Pain19
  • Inflammation13
  • No Clinical Signs, Symptoms or Conditions11
  • Erosion6
  • Swelling/ Edema3
  • Purulent Discharge2

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the implant at tooth site #16 was removed due to peri-implantitis, with associated bone loss and infection.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 50769474 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the implant at tooth site #26 was removed due to peri-implantitis associated with significant bone loss.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 50769120 ↗
  • InjuryAug 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported the implant at tooth site 7 was removed due to infection.

    Patient problems: Unspecified Infection

    MAUDE report 50264223 ↗
  • InjuryJul 31, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that patient returned due to pus around the implant and severe bone loss. Implant was removed due to peri-implantitis. Noted Pain, inflammation and edema. Tooth site 9.

    Patient problems: Osteolysis; Purulent Discharge; Unspecified Infection

    MAUDE report 50096386 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #17 WAS REMOVED DUE TO INFECTION. SYMPTOMS AS A RESULT OF THE EVENT: PAIN, INFLAMMATION AND ABSCESS.

    Patient problems: Inflammation; Pain; Unspecified Infection

    MAUDE report 25322838 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #27 WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 25322502 ↗
  • InjuryMay 27, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTS AT TOOTH SITES #34 AND #36 WERE REMOVED DUE TO PERI-IMPLANTITIS. THE PROCEDURE WAS NOT COMPLETED WITH OTHER IMPLANTS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 25305833 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #7 WAS REMOVED DUE TO BONE LOSS AND INFECTION. PATIENT CAME IN BECAUSE CROWN WAS LOOSE, DOCTOR REMOVED AND CLEANED OUT INFECTION AND PACKED BONE TO BUILD UP BONE. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 25277868 ↗
  • InjuryMay 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS SWELLING AND DRAINAGE AROUND IMPLANT AT TOOTH SITE #23. THE IMPLANT WAS REMOVED DUE TO INFECTION. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.

    Patient problems: Inflammation; Purulent Discharge; Unspecified Infection

    MAUDE report 25086802 ↗
  • InjuryApr 30, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE IMPLANT AT TOOTH SITE #32 WAS REMOVED DUE TO AN INFECTION.BONE TYPE: II.

    Patient problems: Unspecified Infection

    MAUDE report 25040684 ↗
  • MalfunctionApr 3, 2026Separation Failure

    IT WAS REPORTED THAT THE MOUNT WAS STUCK ON THE IMPLANT AT TOOTH SITE #23. THE PROCEDURE WAS COMPLETED USING ANOTHER IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781195 ↗
  • InjuryMar 13, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT IMPLANT WAS REMOVED DUE TO BONE LOSS. THIS WAS AN IMMEDIATE IMPLANT (B)(6) PLACED AFTER EXT. FOR #11. HEALING WAS UNEVENTFUL, 3MO PA LOOKED ACCEPTABLE, I CHECKED TORQUE OF ENCODE HA PRIOR TO RELEASING PATIENT TO GP FOR RESTO. PATIENT REPORTED PAIN WHEN TORQUE FORCE APPLIED TO H. I REMOVED HEALING ABUTMENT AND NECROTIC BONE SEEN AT CORONAL 1/3 OF IMPLANT, IMPLANT REMOVED WITH HAND TORQUE WRENCH WITH LITTLE RESISTANCE. PROCEDURE COMPLETED USING ANOTHER IMPLANT, 4.1X11.5 IMPLANT REPLACED IMMEDIATELY FOLLOWING DEBRIDEMENT OF NECROTIC BONE 45 N-CM INSERTION TORQUE. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Pain

    MAUDE report 24588746 ↗
  • InjuryFeb 25, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT IMPLANT WAS REMOVED DUE TO INFECTION AND BONE LOSS. PATIENT PRESENTED WITH PAIN, WE TAKE XRAY AND FOUND INFECTION AND BONE LOSS, IMPLANT WITH GRADE III MOBILITY. TOOTH SITE # 32.SYMPTOMS AS A RESULT OF THE EVENT: PAIN, INFLAMMATION.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 24446203 ↗
  • MalfunctionFeb 24, 2026Separation Failure

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434386 ↗
  • InjuryFeb 13, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24333040 ↗
  • InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #22 WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 24327387 ↗
  • InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24283359 ↗
  • InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24283275 ↗
  • InjuryFeb 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THERE WAS CRESTAL BONE LOSS AROUND THE IMPLANT AT TOOTH SITE #8. THE IMPLANT WAS REMOVED.

    Patient problems: Osteolysis

    MAUDE report 24267351 ↗
  • InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTS AT TOOTH SITES #14 AND #15 WERE REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS.

    Patient problems: Erosion; Osteolysis; Unspecified Infection

    MAUDE report 24151420 ↗
  • InjuryJan 14, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #21 WAS REMOVED DUE TO PERI-IMPLANTITIS. NOTED FAILURE AT STAGE II. ADDITIONAL APPOINTMENT REQUIRED TO PLACE NEW IMPLANT FUTURE. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 24066259 ↗
  • InjuryJan 7, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Pain

    MAUDE report 23991845 ↗
  • InjuryJan 6, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #10 WAS REMOVED DUE TO INFECTION. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION AND PAIN.

    Patient problems: Inflammation; Pain; Unspecified Infection

    MAUDE report 23981926 ↗
  • InjuryDec 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTS AT TOOTH SITES #14, #23, #34, AND #45 WERE REMOVED DUE TO BONE LOSS.

    Patient problems: Osteolysis

    MAUDE report 23797625 ↗
  • InjuryDec 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 23720325 ↗

Other Implant, Endosseous, Root-Form devices