Skip to content

TSX Implants TSX54B8Recalls & reported problems

ZimVie US Corp LLC · TSX™ IMPLANT, 5.4MMD, 8MML · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This TSX short dental implant by ZimVie US Corp LLC is a screw-shaped implant used to replace missing teeth. The 31 adverse event reports describe implants that loosened and had to be removed, mostly because of bone loss around the implant and infection, with pain, swelling and numbness also mentioned, plus two reports of bone fracture. No deaths were reported and 31 injuries were. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
31
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 00844868045267), received between Oct 12, 2023 and May 5, 2026 — classified as 31 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 4 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem31
  • Loss of Osseointegration1

Patient problems

  • Osteolysis20
  • Unspecified Infection15
  • Pain6
  • Inflammation4
  • Swelling/ Edema4
  • Bone Fracture(s)2
  • Sinus Perforation2
  • Discomfort1

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 WAS REMOVED DUE TO BONE LOSS. DURING FOLLOW UP APPOINTMENT NOTICED MOBILITY OF IMPLANT. REMOVED IMPLANT AND WILL GIVE TIME TO HEAL, WILL HAVE FOLLOW UP APPOINTMENT FOR NEW IMPLANT PLACEMENT.

    Patient problems: Osteolysis

    MAUDE report 25081194 ↗
  • InjuryApr 1, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #18 WAS REMOVED DUE TO PERI-IMPLANTITIS. A DENTAL IMPLANT SITE #18 WAS PLACED USING GUIDED PROTOCOL AND FLAPLESS ENTRY METHOD. 5 MONTHS OF HEALING ALLOWED AND INTEGRATION ACHIEVED. SCREW RETAINED CROWN WAS INSERTED AND PT STARTED USING TOOTH WITH NO PROBLEMS. ABOUT 15 MONTHS LATER, PT PRESENTED WITH PAIN AND CBCT SHOWED BONE LOSS AROUND IMPLANT WITH IMPLANT BODY LOOSE. IMPLANT WAS REMOVED AND SITE CURETTAGE/GRAFTED. SYMPTOMS AS A RESULT OF THE EVENT: PAIN & EDEMA.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 24758555 ↗
  • InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT IMPLANT WAS REMOVED DUE TO BONE LOSS. PATIENT CAME IN FOR CHECK UP DUE TO INSTABILITY, BONE LOSS WAS PRESENT. TOOTH NUMBER: 14.

    Patient problems: Osteolysis

    MAUDE report 24747820 ↗
  • InjuryJan 7, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 23991450 ↗
  • InjuryDec 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #14 WAS REMOVED DUE TO PERI-IMPLANTITIS. PATIENT REPORTED NOTICING HER IMPLANT WAS LOOSE RECENTLY. UPON EXAM AND RADIOGRAPH INTERPRETATION, NOTED RADIOLUCENCY AROUND IMPLANT BODY. PATIENT DID NOT HAVE OCCLUSAL CONTACTS WITH THE OPPOSING DENTITION. PRESCRIBED XANAX PRIOR TO IMPLANT REMOVAL PER PATIENT REQUEST. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION AND SWELLING.

    Patient problems: Inflammation; Osteolysis; Unspecified Infection

    MAUDE report 23866645 ↗
  • InjuryDec 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #30 FAILED ONE (1) YEAR AFTER LOADING AND WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS, PAIN, EDEMA & INFLAMMATION.

    Patient problems: Osteolysis; Swelling/ Edema; Unspecified Infection

    MAUDE report 23829625 ↗
  • InjuryNov 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT WAS REMOVED DUE TO BUCCAL BONE FRACTURE THAT LED TO LACK OF PRIMARY STABILITY. SURGICAL INTERVENTION WAS NOT REQUIRED TO PREVENT PERMANENT DAMAGE. THE PROCEDURE WAS NOT COMPLETED, SITE HAD TO BE RE-GRAFTED AND THE PATIENT WILL RETURN TO A FUTURE APPOINTMENT TO PLACE A NEW IMPLANT.

    Patient problems: Bone Fracture(s)

    MAUDE report 23451800 ↗
  • InjuryOct 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #18 WAS REMOVED DUE TO NERVE INJURY. PATIENT NOTED PARESTHESIA AT IMPLANT SITE. IMPLANT MOBILITY WAS NOTED. THE IMPLANT WAS REMOVED USING TORQUE WRENCH AND BONE GRAFT WAS PLACED. AN ADDITIONAL APPOINTMENT WAS ANTICIPATED TO REPLACE THE IMPLANT. SYMPTOMS AS A RESULT OF THE EVENT: EDEMA AND PARESTHESIA.

    Patient problems: Nerve Damage; Paresthesia; Swelling/ Edema

    MAUDE report 23429825 ↗
  • InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #37 WAS REMOVED DUE TO BONE LOSS.

    Patient problems: Osteolysis

    MAUDE report 23360575 ↗
  • InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #35 WAS REMOVED DUE TO BONE LOSS.

    Patient problems: Osteolysis

    MAUDE report 23360425 ↗
  • InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #14 WAS REMOVED DUE TO AN INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 23344490 ↗
  • InjuryJul 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #31 WAS REMOVED DUE TO BONE LOSS. DURING PRE-RESTORATIVE CHECK, RADIOGRAPHIC EVIDENCE OF SEVERE BONE LOSS PROGRESSION.

    Patient problems: Osteolysis

    MAUDE report 22440296 ↗
  • InjuryJun 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Sinus Perforation

    MAUDE report 22328501 ↗
  • InjuryJun 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAS BEEN EXPERIENCING PAIN SINCE THE PLACEMENT OF THE IMPLANTS AT TOOTH SITES #26 AND #30. THE IMPLANTS WERE REMOVED DUE TO PAIN.

    Patient problems: Pain

    MAUDE report 22158738 ↗
  • InjuryMay 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #30 WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.

    Patient problems: Inflammation; Osteolysis; Unspecified Infection

    MAUDE report 22005666 ↗
  • InjuryMay 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #3 FAILED FOUR (4) MONTHS AFTER PLACEMENT AND WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 21961966 ↗
  • InjuryMar 13, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 21599872 ↗
  • InjuryJan 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED PAIN AFTER RESTORATION NOTED, DISTAL POCKET ON IMPLANT. TOOTH NUMBER 31. IMPLANT WAS REMOVED. SYMPTOMS AS A RESULT OF THE EVENT: PAIN, EDEMA AND INFLAMMATION.

    Patient problems: Inflammation; Pain; Swelling/ Edema

    MAUDE report 21178181 ↗
  • InjuryDec 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT IMPLANT WAS REMOVED DUE TO BONE LOSS. TOOTH 3.

    Patient problems: Osteolysis

    MAUDE report 20906723 ↗
  • InjuryDec 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Unspecified Infection

    MAUDE report 20906372 ↗
  • InjuryNov 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #2 WAS REMOVED DUE TO BONE LOSS. THE IMPLANT WAS REMOVED & REPLACED THE SAME DAY.

    Patient problems: Osteolysis

    MAUDE report 20763460 ↗
  • InjuryNov 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE 26 WAS REMOVED DUE TO PERI-IMPLANTITIS. IT WAS NOTICED DURING A CHECK UP 2 MONTHS AFTER PLACING THE CROWN.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 20722090 ↗
  • InjuryOct 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE # 19 WAS REMOVED DUE TO PERI-IMPLANTITIS. DURING FOLLOW UP PATIENT WAS COMPLAINING OF SEVERE PAIN AND GUMS WERE EXTREMELY IRRITATED. IMPLANT WAS REMOVED USING IMPLANT DRIVER, BONE GRAFT WITH PRP/PRF WAS PLACED.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 20438598 ↗
  • InjuryOct 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT FAILED TO INTEGRATE AND WAS REMOVED DUE TO PERI-IMPLANTITIS. TOOTH LOCATION IS UNKNOWN.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 20374956 ↗
  • InjurySep 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AT FINAL CHECK, THE PATIENT WANTED TO WAIT TO SEE IF LONGER HEALING WOULD RESOLVE DISCOMFORT AROUND THE IMPLANT AT TOOTH LOCATION #14. HOWEVER DURING A TORQUE TEST, THE PA SHOWED BONE LOSS AND TENDERNESS WITH PALPATION. THE IMPLANT WAS THEN REMOVED.

    Patient problems: Discomfort; Osteolysis

    MAUDE report 20332991 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page