T3 T3PT5485Recalls & reported problems

ZimVie US Corp LLC · T3 PRO Tapered Implant 5/4mm (D) X 8.5mm (L) · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This is a ZimVie T3 PRO tapered dental implant, a small screw-shaped post placed in the jaw to anchor a replacement tooth. The 37 reports mostly describe implants that loosened or broke down in the bone and had to be removed, along with pain, swelling, and infection around the implant site. These reports include 35 injuries and no deaths. There have been no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
38
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00844868046806), received between Dec 6, 2022 and Sep 28, 2026 — classified as 38 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem36
  • Mechanical Problem3
  • Patient-Device Incompatibility1

Patient problems

  • Osteolysis25
  • Unspecified Infection14
  • Pain12
  • Inflammation4
  • Nerve Damage4
  • No Clinical Signs, Symptoms or Conditions4
  • Swelling/ Edema3
  • Paresthesia2

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported peri-implantitis, when switching the cover screw for an encode healing abutment for #19,20 implant, fair amount of granulation was found around both implants and they were loose. Implant had to be removed.This event refers to the peri-implantitis in tooth site 19.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 50914383 ↗
  • InjurySep 7, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Osteolysis

    MAUDE report 50650450 ↗
  • InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the implant at tooth #36 was removed due to loss of osseointegration and peri-implantitis. The patient reported pain.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 50034422 ↗
  • InjuryMay 8, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #30 LOST INTEGRATION AND WAS REMOVED DUE TO SEVERE LOCALIZED PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 25125656 ↗
  • InjuryMay 7, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #18 WAS REMOVED DUE TO BONE LOSS. TOOTH NUMBER 18. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Pain

    MAUDE report 25116229 ↗
  • MalfunctionApr 9, 2026Mechanical Problem

    THE DOCTOR REPORTS THAT THE IMPLANT AT TOOTH SITE 46 HAS A DAMAGED HEXAGON AND WAS REMOVED. NO NEGATIVE IMPACT ON THE PATIENT¿S HEALTH WAS REPORTED AS A CONSEQUENCE OF THE EVENT. THE PROCEDURE WAS COMPLETED USING ANOTHER IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24829853 ↗
  • InjuryApr 8, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS SENT TO OUR OFFICE FROM THE GENERAL DENTIST BECAUSE THE PATIENT FELT LIKE THE CROWN TO THE IMPLANT AT TOOTH SITE #14 NEEDED TO BE TIGHTEN. DURING OUR EXAMINATION, THE PATIENT HAD RADIOLUCENCY AROUND THE IMPLANT. THE IMPLANT HAD FAILED AND WAS REMOVED DUE TO PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 24815640 ↗
  • InjuryMar 19, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #14 WAS REMOVED DUE TO BONE LOSS, AND TO PREVENT FURTHER BONE LOSS FROM OCCURRING. THE IMPLANT WAS PLACED WITH AN INDIRECT SINUS LIFT PROCEDURE. TWO WEEKS LATER, THE AREA WAS SWOLLEN AND PAINFUL. BONE LOSS WAS CONFIRMED VIA X-RAY IMAGING AND THE IMPLANT WAS REMOVED. AN ADDITIONAL PROCEDURE IS ANTICIPATED TO REPLACE THE IMPLANT IN SEVERAL MONTHS. SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION AND SWELLING.

    Patient problems: Inflammation; Osteolysis; Swelling/ Edema

    MAUDE report 24641400 ↗
  • InjuryFeb 20, 2026Patient-Device Incompatibility

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Hypersensitivity/Allergic reaction; Osteolysis; Swelling/ Edema

    MAUDE report 24405788 ↗
  • InjuryDec 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #16 WAS REMOVED DUE TO BONE LOSS.

    Patient problems: Osteolysis

    MAUDE report 23867083 ↗
  • InjuryDec 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Sinus Perforation

    MAUDE report 23726260 ↗
  • InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #46 WAS REMOVED DUE TO PERI-IMPLANTITIS. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 23644086 ↗
  • InjuryOct 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 23306433 ↗
  • InjuryJul 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH LOCATION #19 FAILED AND WAS REMOVED DUE TO BONE LOSS.

    Patient problems: Osteolysis

    MAUDE report 22506255 ↗
  • InjuryMay 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE # 13 WAS REMOVED DUE TO PERI-IMPLANTITIS. IMPLANT 13 BECAME TENDER AND MOBILE, WITH BONE LOSS. SYMPTOMS AS A RESULT OF THE EVENT: ABSCESS, INFLAMMATION & PAIN.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 22027837 ↗
  • InjuryMar 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #14 WAS REMOVED DUE TO BONE LOSS. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Osteolysis; Pain

    MAUDE report 21529392 ↗
  • InjuryFeb 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #29 WAS REMOVED DUE TO AN INFECTION WHICH WAS DISCOVERED DURING RADIOLUCENCY. AFTER IMPLANT REMOVAL THE CUSTOMER WAS TREATED WITH ANTIBIOTICS. SYMPTOMS AS A RESULT OF THE EVENT; ABSCESS.

    Patient problems: Erosion; Unspecified Infection

    MAUDE report 21471515 ↗
  • InjuryFeb 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #14 WAS REMOVED DUE TO SEVERE LOCALIZED PERI-IMPLANTITIS.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 21439970 ↗
  • InjuryFeb 18, 2025Mechanical Problem; Mechanical Problem

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21394621 ↗
  • InjuryJan 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #30 HAD AN INFECTION AND HAD TO BE REMOVED.

    Patient problems: Unspecified Infection

    MAUDE report 21223539 ↗
  • InjuryNov 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED NERVE DAMAGE AT TOOTH SITE 46 OBSERVED DURING CLINICAL PROCEDURE. A SHORTER IMPLANT WAS GOING TO BE PLACED DUE TO NERVE PROXIMITY, NEVERTHELESS, THE PATIENT REPORTED HIGH PAIN AND AN ADDITIONAL APPOINTMENT WAS REQUIRED. SYMPTOMS SUFFERED BY THE PATIENT: PARESTHESIA AND PAIN.

    Patient problems: Nerve Damage; Pain; Paresthesia

    MAUDE report 20703987 ↗
  • InjuryNov 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20603998 ↗
  • InjuryOct 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #14 WAS REMOVED DUE TO BONE LOSS. MOBILITY OF #14 IMPLANT AND PAIN WHEN TORQUING OF THE IMPLANT. NO CLEAR REASON FOR THE FAILURE. HAD TO REMOVE AND DEBRIDE IMPLANT SOCKET.

    Patient problems: Osteolysis; Pain

    MAUDE report 20464673 ↗
  • InjuryOct 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DOCTOR REPORTED THAT IMPLANT AT TOOTH SITE 36 WAS REMOVED DUE TO PERI-IMPLANTITIS. SUPPURATION, EDEMA AND PAIN WERE REPORTED AS A RESULT OF THE EVENT. PATIENT WILL RETURN TO PLACE A NEW IMPLANT.

    Patient problems: Osteolysis; Pain; Unspecified Infection

    MAUDE report 20344527 ↗
  • InjurySep 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #4 WAS REMOVED DUE TO INFECTION. INFECTION IN BONE CAUSED IMPLANT TO FAIL. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 20321169 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →