Skip to content

Phoenix 2.2mm x 130cm 6F Deflecting Atherectomy Catheter PD22130Recalls & reported problems

Volcano Atheromed, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845225002985), received between Feb 18, 2021 and Feb 18, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2021Material Separation; Material Separation

    THIS CASE WAS REVIEWED AND INVESTIGATED ACCORDING TO THE MANUFACTURERS POLICY. IT WAS REPORTED AFTER A PLANNED THERAPEUTIC PERIPHERAL PROCEDURE, THE SHAFT OF THE DEVICE APPEARED TO HAVE A FORMED A SMALL RIP (FRACTURE), ALLOWING FLUID TO ESCAPE. THE DEVICE WAS REMOVED FROM THE BODY OVER THE WIRE, AFTER THE DEVICE SEEMED TO BIND UP AND STOP WORKING. ADDITIONAL INFORMATION PROVIDED, THE DEVICE HAD BEEN INSERTED INTO THE PATIENT ONCE. THERE WERE NO ISSUES BEING REMOVED. THE DEVICE WAS RE-FLUSHED AFTER REMOVAL AND SALINE/BLOOD WAS BEING EJECTED FROM THE SHAFT OF THE DEVICE. NOTHING APPEARED TO BE PROTRUDING. A SECOND OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE WAS DISCARDED AT THE FACILITY AND NOT RETURNED FOR ANALYSIS. THE MANUFACTURING DOCUMENTATION FOR THIS DEVICE WAS REVIEWED AND THE DEVICE MET ALL QUALITY AND MANUFACTURING RELEASE CRITERIA. THIS PRODUCT PROBLEM IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION AS IT COULD NOT BE DETERMINED WHERE OR WHEN THE ALLEGED SEPARATION OCCURRED. THERE IS A POTENTIAL FOR HARM IF THE ALLEGED SEPARATION WERE TO RECUR.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 11346812 ↗

Other Catheter, Peripheral, Atherectomy devices

All Catheter, Peripheral, Atherectomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page