Skip to content

CORE Mobile System Medical Equipment 400-0100.07Recalls & reported problems

Philips Image Guided Therapy Corporation · CORE Mobile System - 240V · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00845225012274), received between Oct 24, 2024 and Aug 13, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

10 other recalls cover devices cleared under the same FDA submission (K123898); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device2
  • No Device Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT A CORE MOBILE SYSTEM WAS USED IN A PROCEDURE. DURING USE, PRESSURE MEASUREMENTS WERE NOT POSSIBLE. THE SYSTEM WAS UNPLUGGED AND REBOOTED, BUT COULD NOT BE RESOLVED. A BREAK WAS OBSERVED ON THE FFR PIMMETTE CABLE; THEREFORE, THE CABLE WAS REPLACED BUT DID NOT FULLY RESOLVE THE ISSUE. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED AND IF THE DEVICE WAS BEING USED IN AN EMERGENT PROCEDURE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE IT IS UNKNOWN IF THE FAILURE OCCURRED DURING AN EMERGENT PROCEDURE; POTENTIAL FOR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22795071 ↗
  • MalfunctionNov 29, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT A CORE MOBILE SYSTEM WAS USED IN A CORONARY DIAGNOSTIC EMERGENT PROCEDURE. DURING PREPARATION FOR USE, THE EQUIPMENT FUSE BURNED OUT AND THE SYSTEM SHUT DOWN. THE CASE WAS COMPLETED WITHOUT THE USE OF IVUS. NO INJURY TO PATIENT OR USER WAS REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED DURING AN EMERGENT CASE DUE TO A BURNT FUSE; RESULTING IN A POTENTIAL FOR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20811141 ↗
  • MalfunctionOct 24, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT A CORE MOBILE SYSTEM WAS USED IN A CORONARY THERAPEUTIC PROCEDURE. DURING USE, THE EQUIPMENT FUSE BURNED OUT AND IVUS PROCEDURE WAS STOPPED. THE CASE WAS COMPLETED WITHOUT THE USE OF IVUS. NO INJURY TO PATIENT OR USER WAS REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE CORE MOBILE FUSE WAS BURNT. THERE IS A POTENTIAL FOR HARM IF IT WERE TO RECUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20529246 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123898. One submission often covers many models, so check the product and lot details in each record.

  • Z-0189-2018 · Nov 3, 2017 · Terminated
    Volcano Imaging System s5i REF 807400001; CORE Mobile Imaging System (120V) REF 400-0100.01; CORE Mobile Imaging System Refurbished REF 400-0100.01-R; CORE Mobile Imaging System (2
  • Z-1813-2016 · Apr 25, 2016 · Terminated
    Volcano s5, s5i, CORE and CORE Mobile systems with software version 3.4 and v3.4 software kits. Cardiology: The Volcano Universal (Core /Core Mobile), S5iz Platform Imaging System
  • Z-2131-2015 · Jun 22, 2015 · Terminated
    Volcano s5/s5i/CORE Intravascular Ultrasound Systems with software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5. Radiology: The
  • Z-2136-2015 · Jun 22, 2015 · Terminated
    Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model CORE Radiology: The Vo
  • Z-2135-2015 · Jun 22, 2015 · Terminated
    Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5ix/s5iz Radiology: T
  • Z-2132-2015 · Jun 22, 2015 · Terminated
    Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5x. Radiology: The Vol
  • Z-2134-2015 · Jun 22, 2015 · Terminated
    Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model s5i Radiology: The Vol
  • Z-2133-2015 · Jun 22, 2015 · Terminated
    Volcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout Technology) and v.3.4 software kits; Model CORE Mobile. Radiology:
  • Z-2044-2015 · Sep 23, 2013 · Terminated
    Volcano s5i/s5ix/s5iz/CORE Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5i -806300-020 ; s5ix/s5iz - 807400-001; CORE - 400-0 I 00.02. Intravas
  • Z-2043-2015 · Sep 23, 2013 · Terminated
    Volcano s5/s5x/CORE Mobile Intravascular Ultrasound Imaging and Pressure System; Product/Part numbers: s5 - 804200-002 and 804200-014; s5x - 807300-001; CORE MOBILE - 400-0100.01.

Other System, Imaging, Pulsed Echo, Ultrasonic devices

All System, Imaging, Pulsed Echo, Ultrasonic devices →

Sources and updates