Skip to content

NA AUTOLUBE-III-IRRRecalls & reported problems

Anspach Effort, The · Pneumatic Foot Control for Irrigation System · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00845384000457), received between Oct 14, 2019 and Jul 23, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Leak/Splash1
  • Vibration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2025Break; Leak/Splash; Break; Leak/Splash

    IT WAS REPORTED THAT THE AUTOLUBE DEVICE HOSE WAS DAMAGED AND WAS LEAKING AIR. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE WAS MISSING COMPONENTS, HAD COMPONENT DAMAGE AND THE HOSE WAS RUPTURED. THE DEVICE ALSO FAILED PRETEST FOR VISUAL ASSESSMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578113 ↗
  • MalfunctionDec 29, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE COMPACT SPEED REDUCER HAD VIBRATION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK AND WAS MALFUNCTIONING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11088100 ↗
  • MalfunctionOct 14, 2019Vibration; Vibration

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE BEARING OF THE CRANIOTOME DEVICE WAS WORN. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR TEMPERATURE VERIFICATION AND VIBRATION ASSESSMENT ( ONLY VIBRATION ASSESSMENT FAILED, AND TEMPERATURE VERIFICATION PASSED). IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK ANYMORE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN THE MONTH OF SEPTEMBER. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9183898 ↗

Other MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices

All MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page