The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00845384000457), received between Oct 14, 2019 and Jul 23, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Leak/Splash1
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 23, 2025Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED THAT THE AUTOLUBE DEVICE HOSE WAS DAMAGED AND WAS LEAKING AIR. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE WAS MISSING COMPONENTS, HAD COMPONENT DAMAGE AND THE HOSE WAS RUPTURED. THE DEVICE ALSO FAILED PRETEST FOR VISUAL ASSESSMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22578113 ↗MalfunctionDec 29, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE COMPACT SPEED REDUCER HAD VIBRATION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK AND WAS MALFUNCTIONING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11088100 ↗MalfunctionOct 14, 2019Vibration; Vibration
IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE BEARING OF THE CRANIOTOME DEVICE WAS WORN. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR TEMPERATURE VERIFICATION AND VIBRATION ASSESSMENT ( ONLY VIBRATION ASSESSMENT FAILED, AND TEMPERATURE VERIFICATION PASSED). IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK ANYMORE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN THE MONTH OF SEPTEMBER. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9183898 ↗
Other MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices
- NA 05.001.201Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED1,902 reports
- NA 530.710Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED1,800 reports
- NA 530.705Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED1,492 reports
- NA 530.715Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED390 reports
- NA 530.730Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED267 reports
- NA 05.001.240Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED262 reports
- KINCISE™ 1000-00-101Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED242 reports
- NA 530.791Synthes GmbH · MOTOR, SURGICAL INSTRUMENT, AC-POWERED84 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page