Skip to content

NA QD11-1RRecalls & reported problems

Anspach Effort, The · 1.4 mm x 12.8 mm Fluted Router. Use with QD11 Attachment. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845384005933), received between Jul 15, 2025 and Jul 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Complete Loss of Power” (1 report).

70 other recalls cover devices cleared under the same FDA submission (K011444); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Loss of Power1
  • Excessive Heating1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2025Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating

    IT WAS REPORTED THAT PRIOR TO SURGERY, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE DID NOT WORK ANYMORE. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HEATING UP, THE BEARING WAS WORN AND WAS VIBRATING. THE DEVICE ALSO FAILED PRETEST FOR VIBRATION ASSESSMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22508277 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K011444. One submission often covers many models, so check the product and lot details in each record.

  • Z-0923-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium
  • Z-0919-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
  • Z-0924-2023 · Dec 7, 2022 · Open, Classified
    Pediatric Craniotome Ref:CRANI-P-G1
  • Z-0920-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium
  • Z-0922-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
  • Z-0917-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and crani
  • Z-0916-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium
  • Z-0918-2023 · Dec 7, 2022 · Open, Classified
    Adult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
  • Z-0921-2023 · Dec 7, 2022 · Open, Classified
    Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
  • Z-1947-2017 · Oct 6, 2016 · Terminated
    Electric Handpiece EM Hand Control Product Usage: Electric system
  • Z-1951-2017 · Oct 6, 2016 · Terminated
    eMax2 Hybrid Hand Control Product Usage: Electric system
  • Z-1954-2017 · Oct 6, 2016 · Terminated
    Foot Control without Switches, SC2100 Product Usage: Electric system
  • Z-2029-2017 · Oct 6, 2016 · Terminated
    5.0cm Short Attachment
  • Z-1931-2017 · Oct 6, 2016 · Terminated
    AL-III W/ male disconnect and 20ft hose Product Usage: Pneumatic system
  • Z-2014-2017 · Oct 6, 2016 · Terminated
    16cm Minimally Invasive Attachment Product Usage: Pneumatic system

…and 55 more.

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →

Sources and updates