Skip to content

NA SC2102Recalls & reported problems

Synthes GmbH · eMax 2 Plus Console, Single Port, No Irrigation · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845384011989), received between May 15, 2025 and May 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Complete Loss of Power” (1 report).

35 other recalls cover devices cleared under the same FDA submission (K080802); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Loss of Power1
  • Device Alarm System1
  • Noise, Audible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2025Device Alarm System; Noise, Audible; Complete Loss of Power; Device Alarm System; Noise, Audible; Complete Loss of Power

    IT WAS REPORTED THAT THE SYSTEM CONSOLE DEVICE WAS NOT WORKING PROPERLY AND WOULD ONLY MAKE A CLICKING NOISE WHEN IN USE. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE DEVICE WAS DISPLAYING AN E8 ERROR CODE (SYSTEM FAULT), AN ERRONEOUS ERROR CODE, AND HAD COMPONENT DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR POWER CORD. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22027675 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080802. One submission often covers many models, so check the product and lot details in each record.

  • Z-2031-2017 · Oct 6, 2016 · Terminated
    Sagittal MICRO-SAW
  • Z-2019-2017 · Oct 6, 2016 · Terminated
    Oscillating MICRO-SAW Product Usage: Pneumatic system
  • Z-1957-2017 · Oct 6, 2016 · Terminated
    System Console, Single Port, no irrigation
  • Z-1955-2017 · Oct 6, 2016 · Terminated
    Console, eMax2 Plus Product Usage: Electric system
  • Z-1956-2017 · Oct 6, 2016 · Terminated
    System Console, Single Port, with irrigation Product Usage: Electric system
  • Z-1949-2017 · Oct 6, 2016 · Terminated
    Electronic Foot Control with Direction or Irrigation Switches Product Usage: Electric system
  • Z-2026-2017 · Oct 6, 2016 · Terminated
    Reciprocating MICRO SAW
  • Z-2001-2017 · Oct 6, 2016 · Terminated
    12.4cm Heavy Duty Long Heavy Short Attachment Product Usage: Pneumatic system
  • Z-2032-2017 · Oct 6, 2016 · Terminated
    20.2cm Heavy Duty Long Heavy Short Attachment
  • Z-1948-2017 · Oct 6, 2016 · Terminated
    Electronic Foot Control with Direction Switch Product Usage: Electric system
  • Z-2015-2017 · Oct 6, 2016 · Terminated
    Sagittal Saw Product Usage: Pneumatic system
  • Z-1952-2017 · Oct 6, 2016 · Terminated
    eMax2 PLUS Product Usage: Electric system
  • Z-0727-2014 · Dec 17, 2013 · Terminated
    Anspach Carbide Cutting Burrs: 1.5MMx7.5MM Cylindrical Carbide Drum part # S-15DRMC-L, 1MM Carbide Fluted ball, Extended part # S-1B-C, 1MM Carbide Fluted ball part # S-SB-C, 2MM C
  • Z-0728-2014 · Dec 17, 2013 · Terminated
    Anspach Carbide Cutting Burrs & Drums: 1.5MM x 7.5MM Cylindrical Carbide Drum part # S-15DRMC Anspach cutting burrs are intended for cutting and shaping bone, including the spine
  • Z-0729-2014 · Dec 17, 2013 · Terminated
    Anspach Carbide Cutting Burrs & Drums: 2MM Carbide Fluted Ball, Extended part # S-2B-C and 7MM Carbide Fluted Ball part # S-7B-C Anspach cutting burrs are intended for cutting and

…and 20 more.

Other MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices

All MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page