Skip to content

NA MA-D20-G1Recalls & reported problems

Synthes GmbH · Minimal Access Angle Driver, G1 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes Minimal Access Angle Driver is a powered surgical handpiece attachment used in minimally invasive procedures. The 68 reports mostly describe the tool heating up or vibrating, and sometimes breaking, sticking, or failing to detach from the handpiece. One report mentions a superficial burn; no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
68
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 00845384016397), received between Nov 15, 2018 and Jun 11, 2025 — classified as 1 injury, 67 malfunctions.

The most frequently reported specific device problem is “Excessive Heating” (25 reports).

11 other recalls cover devices cleared under the same FDA submission (K133604); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Excessive Heating25
  • Vibration23
  • Overheating of Device16
  • Detachment of Device or Device Component4
  • Appropriate Device Problem Term/Code Not Available3
  • Complete Loss of Power3
  • Fail-Safe Did Not Operate3
  • Device-Device Incompatibility2

Patient problems

  • No Clinical Signs, Symptoms or Conditions33
  • No Consequences Or Impact To Patient23
  • No Known Impact Or Consequence To Patient11
  • Superficial (First Degree) Burn1

Recent adverse event reports

The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2025Overheating of Device; Device-Device Incompatibility; Overheating of Device; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DRIVER DEVICE WOULD NOT DISENGAGE FROM THE HANDPIECE. DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD HEAT AND FAILED PRE-TEST FOR TEMPERATURE ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22186695 ↗
  • MalfunctionMay 21, 2025Appropriate Device Problem Term/Code Not Available; Premature Separation; Appropriate Device Problem Term/Code Not Available; Premature Separation

    IT WAS REPORTED THAT THE ANGLED DRIVER ATTACHMENT DEVICE WAS BROKEN. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE DEVICE HAD UNINTENDED/UNEXPECTED DISENGAGEMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22064470 ↗
  • MalfunctionAug 9, 2024Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE MINIMAL ACCESS DRIVER DEVICE WAS OVERHEATING. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19947540 ↗
  • MalfunctionJul 17, 2024Mechanical Problem; Fail-Safe Did Not Operate; Mechanical Problem; Fail-Safe Did Not Operate

    IT WAS REPORTED FROM CANADA THAT THE LOCKING MECHANISM ON THE ATTACHMENT DEVICE WAS STIFF. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS OBSERVED THAT THE DEVICE RAN IN LOCKED POSITION. IT WAS FURTHER DETERMINED THAT THE DEVICE HAD CORROSION/RUSTING/PITTING. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CUTTER BRAKE ASSESSMENT. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19760948 ↗
  • MalfunctionAug 18, 2023Complete Loss of Power; Difficult or Delayed Separation; Fail-Safe Did Not Operate; Complete Loss of Power; Difficult or Delayed Separation; Fail-Safe Did Not Operate

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE OPERATED INTERMITTENTLY, IT WOULD SPIN AND THEN STOP. IT WAS ALSO NOTED THAT WHEN PULLING OUT THE CUTTER DEVICES THE DEVICE WOULD FEEL STUCK AND WAS DIFFICULT TO REMOVE. ACCORDING TO THE REPORTER, FOUR MOTOR DEVICES WERE TESTED WITH THE ATTACHMENT DEVICE WITH THE SAME RESULT. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE RUN IN THE LOCKED POSITION. THE DEVICE ALSO FAILED PRETESTS FOR CUTTER BRAKE ASSESSMENT. IT WAS UNKNOWN IF THERE WAS A DELAY IN THE SURGICAL PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17573547 ↗
  • MalfunctionMay 2, 2023Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS OVERHEATING WHILE IN USE. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16852795 ↗
  • MalfunctionMar 31, 2023Excessive Heating; Difficult or Delayed Separation; Excessive Heating; Difficult or Delayed Separation

    IT WAS REPORTED FROM CANADA THAT THE ATTACHMENT DEVICE WAS HEATING UP AND WAS DIFFICULT TO REMOVE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16659094 ↗
  • MalfunctionMar 10, 2023Appropriate Device Problem Term/Code Not Available; Excessive Heating; Appropriate Device Problem Term/Code Not Available; Excessive Heating

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE MINIMAL ACCESS DRIVE DEVICE WAS NOT WORKING PROPERLY AND WAS HEATING UP. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16522119 ↗
  • MalfunctionMar 10, 2023Appropriate Device Problem Term/Code Not Available; Excessive Heating; Appropriate Device Problem Term/Code Not Available; Excessive Heating

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE MINIMAL ACCESS DRIVE DEVICE WAS NOT WORKING PROPERLY AND WAS HEATING UP. IT WAS REPORTED THAT THERE WAS NO DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16522066 ↗
  • MalfunctionFeb 10, 2023Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS HEATING UP. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16356443 ↗
  • MalfunctionJan 31, 2023Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING TESTING, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS GETTING HOT WHILE RUNNING. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16271171 ↗
  • MalfunctionJan 30, 2023Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING TESTING, IT WAS OBSERVED THAT THE MINIMAL ACCESS DRIVE DEVICE WAS GETTING HOT WHEN RUNNING. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16262818 ↗
  • MalfunctionJan 26, 2023Excessive Heating; Excessive Heating

    THIS IS REPORT 1 OF 2 OF THE SAME EVENT: IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE AND BEARING SLEEVE DEVICE EXPERIENCED EXCESSIVE FRICTION AND HEATED UP WHILE IN USE. IT WAS REPORTED THAT THE DEVICES WERE BEING USED TOGETHER WHEN THE EVENT OCCURRED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16245833 ↗
  • MalfunctionJan 18, 2023Excessive Heating; Excessive Heating

    THIS IS REPORT 1 OF 2 OF THE SAME EVENT IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE MOTOR DEVICE AND MINIMAL ACCESS DRIVER DEVICES HEATED UP. THERE WERE NO REPORTS OF DELAYS IN THE SURGICAL PROCEDURE. IT WAS UNKNOWN IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16199054 ↗
  • MalfunctionDec 14, 2022Excessive Heating; Excessive Heating

    IT WAS REPORTED FROM CANADA THAT THE MINIMAL ACCESS DRIVER DEVICE WAS HEATING UP DURING USE. IT WAS NOT REPORTED IF THE EVENT OCCURRED DURING A SURGICAL PROCEDURE. IT WAS NOT REPORTED IF THERE WAS A DELAY TO A PLANNED PROCEDURE. IT WAS NOT REPORTED IF THERE WAS A SPARE DEVICE AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15982411 ↗
  • MalfunctionOct 11, 2022Device Slipped; Premature Separation; Device Slipped; Premature Separation

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE BEARINGS OF THE ATTACHMENT DEVICE WERE LOOSE, AND THE DRILL BIT DEVICE WAS SLIPPING WHILE IN USE. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15579154 ↗
  • MalfunctionAug 25, 2022Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE MINIMAL ACCESS DRIVE DEVICE WAS GETTING HOT. THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE. IT WAS UNKNOWN IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15294871 ↗
  • MalfunctionJul 21, 2022Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE MINIMAL ACCESS ATTACHMENT DEVICE WAS GETTING HOT. IT WAS UNKNOWN IF THERE WERE ANY DELAYS IN THE PROCEDURE DUE TO THE EVENT. IT WAS NOT REPORTED IF A SPARE DEVOICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15078130 ↗
  • MalfunctionJun 24, 2022Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT THE ATTACHMENT DEVICE WAS OVERHEATING. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14797980 ↗
  • MalfunctionJun 23, 2022Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT THE ATTACHMENT DEVICE WAS OVERHEATING. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14787631 ↗
  • MalfunctionJun 15, 2022Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING PRE-TESTING, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE OVERHEATED. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14696419 ↗
  • MalfunctionJun 8, 2022Excessive Heating; Excessive Heating

    THIS IS REPORT 2 OF 2 OF THE SAME EVENT: IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE MOTOR DEVICE AND ATTACHMENT DEVICE WERE INTERMITTENTLY HEATING UP WHILE BEING USED TOGETHER. THE SURGEON FURTHER REPORTED THAT THEY COULD NOT DETERMINE WHICH DEVICE WAS HEATING UP. IT WAS REPORTED THAT THE MOTOR DEVICE WAS ALSO MAKING A STRANGE NOISE. IT WAS REPORTED THAT THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE. IT WAS NOT REPORTED IF SPARE DEVICES WERE AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14635123 ↗
  • MalfunctionMay 6, 2022Vibration; Noise, Audible; Excessive Heating; Vibration; Noise, Audible; Excessive Heating

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE BECAME HOT, VIBRATED AND MADE NOISE. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14305242 ↗
  • MalfunctionFeb 28, 2022Mechanical Jam; Complete Loss of Power; Excessive Heating; Mechanical Jam; Complete Loss of Power; Excessive Heating

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE MINIMAL ACCESS DRIVER DEVICE WAS OVERHEATING AND THE COLOR BAND WAS CHIPPING. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT AND TEMPERATURE ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS STUCK AND DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13619886 ↗
  • MalfunctionDec 3, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PRE-SURGERY TESTING, IT WAS DISCOVERED THAT THE TIP OF THE ATTACHMENT DEVICE BROKE OFF. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12924076 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133604. One submission often covers many models, so check the product and lot details in each record.

  • Z-0920-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium
  • Z-0923-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium
  • Z-0919-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
  • Z-0924-2023 · Dec 7, 2022 · Open, Classified
    Pediatric Craniotome Ref:CRANI-P-G1
  • Z-0918-2023 · Dec 7, 2022 · Open, Classified
    Adult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
  • Z-0917-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and crani
  • Z-0922-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
  • Z-0916-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium
  • Z-0921-2023 · Dec 7, 2022 · Open, Classified
    Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
  • Z-2253-2016 · Jun 15, 2016 · Terminated
    Anspach G1 High Speed Electric Handpiece Air Cooled (eG1A) Use for cutting and shaping bone including bones of the cranium and spine.
  • Z-1037-2015 · Dec 10, 2014 · Terminated
    16cm Minimally Invasive Attachment. For use with Motor Systems for cutting and shaping bones, including the spine and cranium. Model MIA16-G1

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page