STREAMLINE™* TL SPINAL FIXATION SYSTEM 01-PA-65-45Recalls & reported problems

Pioneer Surgical Technology, Inc. · PEDICLE SCREW, POLYAXIAL · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00846468005658), received between May 31, 2018 and Mar 7, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Improper or Incorrect Procedure or Method2
  • Off-Label Use2
  • Disassembly1

Patient problems

  • Failure of Implant2
  • Intraoperative Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 7, 2024Break; Off-Label Use; Improper or Incorrect Procedure or Method; Break; Off-Label Use; Improper or Incorrect Procedure or Method

    OUR OEM DISTRIBUTION REP STATED "XTANT MEDICAL HAS RECEIVED A PRODUCT COMPLAINT FROM THE STREAMLINE TL SPINAL FIXATION SYSTEM, IN WHICH IT WAS REPORTED THAT TWO PEDICLE SCREW IMPLANTS (01-PA-65-45, LOTS: 381268 AND 380862) WERE IDENTIFIED AS BROKEN, WHICH REQUIRED A REVISION PROCEDURE. XTANT MEDICAL HAS INITIATED COMPLAINT: (B)(4) ON THE ISSUE. BELOW IS RELEVANT INFORMATION PROVIDED BY THE COMPLAINT SOURCE, PLEASE LET ME KNOW IF YOU HAVE ANY QUESTIONS OR CONCERNS, APPRECIATE THE ASSISTANCE. COMMENTARY PROVIDED: PATIENT WAS IMPLANTED ON (B)(6) 2023, ON (B)(6) 2023, PATIENT CAME IN AND REPORTS 95% IMPROVED AND WAS PLEASED WITH SURGERY RESULT. ON APPROXIMATELY ON (B)(6) 2024, A STRANGE SQUEAKING NOISE BEGAN. ON (B)(6) 2024, CT SCAN SHOWS VERTICAL NONDISPLACED FRACTURES THROUGH THE BILATERAL S1 PEDICLE SCREWS AT BOTH OSSEOUS ENTRY POINTS. INDEX PROCEDURE PERFORMED ON (B)(6) 2023. REVISION PROCEDURE PERFORMED ON (B)(6) 2024".

    Patient problems: Failure of Implant

    MAUDE report 18860013 ↗
  • InjuryMar 7, 2024Break; Off-Label Use; Improper or Incorrect Procedure or Method; Break; Off-Label Use; Improper or Incorrect Procedure or Method

    OUR OEM DISTRIBUTION REP STATED "XTANT MEDICAL HAS RECEIVED A PRODUCT COMPLAINT FROM THE STREAMLINE TL SPINAL FIXATION SYSTEM, IN WHICH IT WAS REPORTED THAT TWO PEDICLE SCREW IMPLANTS (01-PA-65-45, LOTS 381268 AND 380862) WERE IDENTIFIED AS BROKEN, WHICH REQUIRED A REVISION PROCEDURE. XTANT MEDICAL HAS INITIATED COMPLAINT (B)(4) ON THE ISSUE. ATTACHED IS THE IMAGING PROVIDED BY THE COMPLAINT SOURCE. BELOW IS RELEVANT INFORMATION PROVIDED BY THE COMPLAINT SOURCE, PLEASE LET ME KNOW IF YOU HAVE ANY QUESTIONS OR CONCERNS, APPRECIATE THE ASSISTANCE. COMMENTARY PROVIDED: PATIENT WAS IMPLANTED (B)(6) 2023, ON (B)(6) 2023 PATIENT CAME IN AND REPORTS 95% IMPROVED AND WAS PLEASED WITH SURGERY RESULT. ON APPROXIMATELY (B)(6) 2024 A STRANGE SQUEAKING NOISE BEGAN. ON (B)(6) 2024 CT SCAN SHOWS VERTICAL NONDISPLACED FRACTURES THROUGH THE BILATERAL S1 PEDICLE SCREWS AT BOTH OSSEOUS ENTRY POINTS. INDEX PROCEDURE PERFORMED (B)(6) 2023, REVISION PROCEDURE PERFORMED (B)(6) 2024".

    Patient problems: Failure of Implant

    MAUDE report 18859552 ↗
  • InjuryMay 31, 2018Disassembly; Disassembly

    A POSTERIOR SPINAL FUSION SURGERY WAS PERFORMED ON THE PATIENT ON (B)(6) 2018. PATIENT STARTED TO HAVE PAIN A FEW WEEKS AFTER THIS INITIAL SURGERY. A CT SCAN WAS COMPLETED FOR THIS PATIENT AND THE SCAN SHOWED THAT THE PEDICLE SCREW HEAD DISASSEMBLED. THE PATIENT WAS REVISED ON (B)(6) 2018 AND THE DISASSEMBLED SCREW WAS REMOVED AND REPLACED.

    Patient problems: Intraoperative Pain

    MAUDE report 7557894 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →