Tornier Latitude 9030375Recalls & reported problems
Tornier, Inc. · HEX SCREW DRIVER BIT · Class 1
All models of Tornier Latitude (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00846832019991), received between May 3, 2021 and Mar 1, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Fracture2
Patient problems
- Foreign Body In Patient2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 1, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OF IN THE SCREW HEAD AND REMAINED IN THE PATIENT. THIS CAUSED A 2 MINUTE DELAY IN SURGERY.
Patient problems: Foreign Body In Patient
MAUDE report 13636174 ↗MalfunctionAug 16, 2021Fracture; Fracture
IT WAS REPORTED THAT DURING A REVISION SURGERY THE TIP OF THE DRIVER BROKE OFF TRYING TO LOOSEN THE ULNA CAP OFF OF THE LOOSE ULNA COMPONENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 12328279 ↗MalfunctionMay 3, 2021Break; Break
IT WAS REPORTED THE DRIVER WAS BEING USED WITH A 10NM TORQUE HANDLE AND THAT THE TIP OF THE DRIVER BROKE OFF IN THE HEAD OF THE SCREW AND WAS COLD WELDED TO THE SCREW HEAD. THE TIP WAS NOT ABLE TO BE RETRIEVED FROM THE PATIENT. THE SURGEON PROCEEDED TO FOLD THE TAB OVER WITH THE PUNCH ON THE ULNAR CAP. NO ADDITIONAL PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 11767762 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page