Tornier Perform® Reversed Glenoid DWJ726Recalls & reported problems
Tornier, Inc. · Peripheral Screw · Class 2 · Implant
All models of Tornier Perform Reversed Glenoid (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00846832091249), received between Jul 18, 2022 and Jul 7, 2026 — classified as 7 injuries.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem6
- Fracture1
- Loss of Osseointegration1
- Osseointegration Problem1
- Unstable1
Patient problems
- Unspecified Infection4
- Loss of Range of Motion3
- Pain3
- Inadequate Osseointegration1
- Joint Laxity1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJul 7, 2026Loss of Osseointegration; Adverse Event Without Identified Device or Use Problem
It was reported that the patient underwent a right primary reverse shoulder arthroplasty. At approximately two-year follow-up, the patient presented with shoulder pain and reduced range of motion. The surgeon noted reported non-compliance with postoperative rehabilitation and a history of potential trauma while climbing in and out of a truck.During revision surgery, the glenoid component was found to be loose and dissociated from the glenoid bone and was easily removed. The humeral component remained well fixed and was retained. The polyethylene insert exhibited eccentric wear, and grey synovial tissue was observed and excised. Copious lavage was performed.New glenoid components were implanted with the addition of synthetic bone graft. A new humeral polyethylene insert was placed. The shoulder was reduced and found to be stable through range of motion.
Patient problems: Loss of Range of Motion; Pain
MAUDE report 25734865 ↗InjuryMay 29, 2026Fracture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT PRIMARY REVERSE SHOULDER ARTHROPLASTY. AT APPROXIMATELY TWO-YEAR FOLLOW-UP, THE PATIENT PRESENTED WITH SHOULDER PAIN AND REDUCED RANGE OF MOTION. THE SURGEON NOTED REPORTED NON-COMPLIANCE WITH POSTOPERATIVE REHABILITATION AND A HISTORY OF POTENTIAL TRAUMA WHILE CLIMBING IN AND OUT OF A TRUCK. DURING REVISION SURGERY, THE GLENOID COMPONENT WAS FOUND TO BE LOOSE AND DISSOCIATED FROM THE GLENOID BONE AND WAS EASILY REMOVED. THE HUMERAL COMPONENT REMAINED WELL FIXED AND WAS RETAINED. THE POLYETHYLENE INSERT EXHIBITED ECCENTRIC WEAR, AND GREY SYNOVIAL TISSUE WAS OBSERVED AND EXCISED. COPIOUS LAVAGE WAS PERFORMED. NEW GLENOID COMPONENTS WERE IMPLANTED WITH THE ADDITION OF SYNTHETIC BONE GRAFT. A NEW HUMERAL POLYETHYLENE INSERT WAS PLACED. THE SHOULDER WAS REDUCED AND FOUND TO BE STABLE THROUGH RANGE OF MOTION.
Patient problems: Loss of Range of Motion; Pain
MAUDE report 25330996 ↗InjurySep 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT; A PATIENT WAS BOOKING IN FOR A REVISION PROCEDURE DUE TO ONGOING PAIN AND STIFFNESS. DURING THE DISSECTION LOTS OF FLUID/ PUSS WAS SEEN AND THE DECISION WAS MADE TO DO A 2 STAGE PROCEDURE TO CLEAR THE INFECTION. ANOTHER REVISION PROCEDURE WILL BE NEEDED AFTER THE INFECTION HAS CLEARED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Loss of Range of Motion; Pain; Unspecified Infection
MAUDE report 23084247 ↗InjuryApr 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION".
Patient problems: Unspecified Infection
MAUDE report 19205365 ↗InjuryMar 5, 2024Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY DUE TO INSTABILITY OF THE JOINT. THE DEVICE WAS REMOVED AND REPLACED WITH NEW HARDWARE.
Patient problems: Joint Laxity; Unspecified Infection
MAUDE report 18842504 ↗InjuryMay 12, 2023Osseointegration Problem; Osseointegration Problem
THERE WAS AN EARLY REVISION OF A PERFORM REVERSED GLENOID IMPLANTED. FLUIDS AND TISSUES HAVE BEEN HARVESTED DURING THE REVISION CASE FOR A SUSPECTED INFECTION. THERE IS NO RESULTS OF THE PATHOLOGY TESTS. ONCE THE REVERSE GLENOID WAS REMOVED, THE SURGEON PACKED THE GLENOID VAULT WITH SOME FEMORAL HEAD GRAFT AND HAS CONVERTED THE STEM TO A HEMI PROSTHESIS. THE STEM WAS NOT REVISED.
Patient problems: Inadequate Osseointegration
MAUDE report 16924375 ↗InjuryJul 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY DUE TO AN INFECTION 12 MONTHS POST-OP. THE METAL DEVICES WERE REMOVED AND A COMPETITOR'S CEMENT SPACER WAS INSERTED.
Patient problems: Unspecified Infection
MAUDE report 15043833 ↗
Other shoulder prosthesis, reverse configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page