Skip to content

Tornier HRS ARS1022801Recalls & reported problems

Tornier, Inc. · Reversed Tray · Class 2 · Implant

All models of Tornier HRS (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00846832094004), received between Jun 10, 2025 and Aug 13, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Difficult to Setup or Prepare1
  • Entrapment of Device1
  • Mechanical Jam1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    IT WAS REPORTED THAT DURING SURGERY, TWO HRS -2 POLYETHYLENE INSERTS WERE UNABLE TO SEAT/LOCK INTO THE HRS CENTERED TRAY DESPITE BEING IMPACTED INTO PLACE. BOTH INSERTS WERE ATTEMPTED IN VIVO AND ON THE BACK TABLE AND DID NOT CONNECT, RESULTING IN THEIR DISPOSAL. THE CENTERED TRAY WAS REMOVED FROM USE, AND A NEW CENTERED TRAY WAS OPENED; HOWEVER, THE ORIGINAL TWO INSERTS AGAIN FAILED TO SEAT. AFTER A SIGNIFICANT DELAY, A THIRD INSERT ¿ A NON-VITAMIN E STANDARD PERFORM +0 POLY ¿ WAS OPENED AND SUCCESSFULLY IMPLANTED WITH AN ECCENTRIC TRAY. THE ECCENTRIC TRAY USED FOR IMPLANTATION HAD ALSO FAILED TO SEAT THE ORIGINAL TWO INSERTS.

    Patient problems: Insufficient Information

    MAUDE report 22794807 ↗
  • MalfunctionJun 10, 2025Entrapment of Device; Mechanical Jam; Entrapment of Device; Mechanical Jam

    THE HRS STEM WAS USED. THE POLY WAS TRAILED AND -2 WAS SELECTED. THE FINAL IMPLANT WAS OPENED. IT WAS A STRUGGLE TO GET THE POLY TO SIT CORRECTLY, SO IT WAS TAKEN OFF WITH A FREER AND TRIED AGAIN. THE IMPLANT SUCCESSFULLY SEATED AND WHEN THE SHOULDER WAS REDUCED, THE JOINT WAS LOOSE. IT WAS DECIDED TO TAKE THE POLY OFF AND RE-TRIAL WITH A 0 POLY. TO GET THE -2 POLY OFF, A PERIPHERAL DRILL WAS USED TO DRILL IN THE CENTER OF THE IMPLANT, THEN A PERIPHERAL SCREW WAS USED TO DRILL THE POLY BUT IT DIDN¿T POP OFF. THEN A 6.5 SCREW WAS USED AND THE SAME THING HAPPENED. NEXT, A 9.5 AND AGAIN THE POLY DIDN¿T POP OFF. AT THIS POINT, AN OSTEOTOME WAS USED TO GET THE POLY OFF AND PIECES OF THE POLY WOULD COME OFF BUT NOT THE WHOLE THING. THE TRAY WITH THE TUNING FORK HAD TO BE TAKEN OFF. THERE WAS ONLY ONE 8 CENTERED HRS TRAY AT THE FACILITY SO THE POLY NEEDED TO BE OFF ON THE BACK TABLE WITH AN OSTEOTOME. THE POLY CRACKED AND ONLY PIECES OF THE POLY CAME OFF. INSTEAD, AN 8 OFFSET TRAY WAS USED SINCE THE 0 TRAY NOW HAD HALF OF A POLY STUCK IN IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22180239 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page