Skip to content

Tornier HRS ARS1023305Recalls & reported problems

Tornier, Inc. · Trial Proximal Body 135° · Class 1

All models of Tornier HRS (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00846832094516), received between Dec 3, 2025 and Jun 16, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Difficult to Setup or Prepare1
  • Difficult to Remove1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 16, 2026Mechanical Jam

    IT WAS REPORTED THAT THE ARS743000 WOULD GET STUCK ON THE PROXIMAL BODY TRIALS AFTER EVERY IMPACTION. ACCORDING TO THE STRYKER REP, THIS MADE THE TRIALLING VERY DIFFICULT (NEED A WAY TO LOOSEN HOW TIGHT THE HANDLE MATES ONTO THE PROXIMAL BODY TRIALS). ADDITIONAL INFORMATION FROM THE STRYKER REP MAY 21: "I TRIED THE HANDLE ON ONE OF OUR LTC PROXIMAL BODIES AND VICE-VERSA, AND IT TURNS OUT THAT THE HANDLE IS NOT THE PROBLEM AT ALL, IT WORKED PERFECTLY. THE PROBLEM IS WITH THE NEW PROXIMAL BODY TRIALS (ARS1023301-ARS1023306). THEY WERE STICKY ON THE OLD HANDLE AS WELL".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25521882 ↗
  • MalfunctionDec 5, 2025Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    IT WAS REPORTED THAT THE HRS PROXIMAL BODY SPACER SCREW MALFUNCTIONED. THE SCREW WAS NOT ADVANCING AND JUST SPINNING IN THE PROXIMAL BODY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23727028 ↗
  • MalfunctionDec 3, 2025Difficult to Remove; Difficult to Remove

    THE NEW PROXIMAL BODIES FOR HRS ARE GETTING STUCK IN THE HANDLES. THE PHYSICIAN HAS TRIED AN OLD REVIVE HANDLE AND ORDERED A BRAND NEW HANDLE. THE PROXIMAL BODIES HAVE GOTTEN STUCK ON MULTIPLE OCCASIONS AND CAUSED THEM TO HAVE TO TAP THE STEM OFF WITH A MALLET INTRA OP. SOMETIMES THE STEM COMES WITH IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23709083 ↗

Other TEMPLATE devices

All TEMPLATE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page