Skip to content

BioHorizons TLXP3809Recalls & reported problems

Biohorizons Implant Systems, Inc · Tapered Internal Plus Implant, 3.8mm x 9mm, 3.0 Platform, Laser-Lok · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This BioHorizons Tapered Internal Plus is an implant placed into bone. Most of the 134 reports describe the implant failing to bond with the bone, and some describe its removal. There are no deaths and 134 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
134
Deaths reported
0
Injuries reported
134
Recalls naming it
0

Summary

The FDA's MAUDE database contains 134 adverse event reports that list this exact device (DI 00847236009601), received between Feb 24, 2024 and Sep 7, 2026 — classified as 134 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (108 reports).

1 other recall cover devices cleared under the same FDA submission (K121787, K223697); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 4 reportsMar 2024: 11 reportsApr 2024: 4 reportsMay 2024: 10 reportsJun 2024: 2 reportsJul 2024: 6 reportsAug 2024: 3 reportsSep 2024: 2 reportsOct 2024: 8 reportsNov 2024: 3 reportsDec 2024: 11 reportsJan 2025: 3 reportsFeb 2025: 4 reportsMar 2025: 1 reportsApr 2025: 10 reportsMay 2025: 5 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 5 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 7 reportsJan 2026: 1 reportsFeb 2026: 5 reportsMar 2026: 5 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 6 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate108
  • Use of Device Problem30
  • Loss of Osseointegration3
  • Fracture1
  • Osseointegration Problem1

Patient problems

  • Insufficient Information134

Recent adverse event reports

The 25 most recent of 134 reports, in the reporter's own words as published by the FDA.

  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Insufficient Information

    MAUDE report 50643176 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Insufficient Information

    MAUDE report 50643174 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    Implant failed to osseointegrate.

    Patient problems: Insufficient Information

    MAUDE report 50280721 ↗
  • InjuryAug 16, 2026Failure to Osseointegrate

    Implant failed to osseointegrate.

    Patient problems: Insufficient Information

    MAUDE report 50268415 ↗
  • InjuryAug 16, 2026Use of Device Problem

    Implant removed due to surgical consideration within 36 hours.

    Patient problems: Insufficient Information

    MAUDE report 50268392 ↗
  • InjuryAug 14, 2026Failure to Osseointegrate

    Implant failed to osseointegrate.

    Patient problems: Insufficient Information

    MAUDE report 50256343 ↗
  • InjuryAug 7, 2026Use of Device Problem

    Implant removed due to surgical consideration within 36 hours.

    Patient problems: Insufficient Information

    MAUDE report 50181748 ↗
  • InjuryAug 5, 2026Failure to Osseointegrate

    Implant failed to osseointegrate.

    Patient problems: Insufficient Information

    MAUDE report 50139950 ↗
  • InjuryJul 12, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 25791503 ↗
  • InjuryJul 10, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE

    Patient problems: Insufficient Information

    MAUDE report 25781274 ↗
  • InjuryJul 9, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 25773710 ↗
  • InjuryMay 26, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 25278672 ↗
  • InjuryMay 13, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 25160508 ↗
  • InjuryApr 30, 2026Use of Device Problem

    IMPLANT REMOVED DUE TO SURGICAL CONSIDERATION WITHIN 36 HOURS.

    Patient problems: Insufficient Information

    MAUDE report 25052052 ↗
  • InjuryApr 30, 2026Use of Device Problem

    IMPLANT REMOVED DUE TO SURGICAL CONSIDERATION WITHIN 36 HOURS.

    Patient problems: Insufficient Information

    MAUDE report 25051063 ↗
  • InjuryMar 22, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24658456 ↗
  • InjuryMar 18, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24630572 ↗
  • InjuryMar 16, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24605538 ↗
  • InjuryMar 15, 2026Use of Device Problem

    IMPLANT REMOVED DUE TO SURGICAL CONSIDERATION WITHIN 36 HOURS.

    Patient problems: Insufficient Information

    MAUDE report 24601893 ↗
  • InjuryMar 9, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24544931 ↗
  • InjuryFeb 25, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24446297 ↗
  • InjuryFeb 24, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24433904 ↗
  • InjuryFeb 23, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24414179 ↗
  • InjuryFeb 19, 2026Failure to Osseointegrate

    IMPLANT FAILED TO OSSEOINTEGRATE.

    Patient problems: Insufficient Information

    MAUDE report 24391759 ↗
  • InjuryFeb 14, 2026Use of Device Problem

    IMPLANT REMOVED DUE TO SURGICAL CONSIDERATION WITHIN 36 HOURS.

    Patient problems: Insufficient Information

    MAUDE report 24340806 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121787, K223697. One submission often covers many models, so check the product and lot details in each record.

  • Z-1111-2025 · Dec 6, 2024 · Open, Classified
    Temporary Titanium Abutments

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page