Skip to content

HyperGlideTM 104-4127Recalls & reported problems

Micro Therapeutics, Inc. · 104-4127 HYPERGLIDE 4X20MM BALLOON SYS · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536005020), received between May 15, 2019 and May 15, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Known Impact Or Consequence To Patient1
  • Paresis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2019Burst Container or Vessel; Burst Container or Vessel

    MEDTRONIC RECEIVED INFORMATION THAT THE HYPERGLIDE BALLOON BURST DURING A PROCEDURE. PRIOR TO THE EVENT, THE PATIENT HAD JUST UNDERGONE A MECHANICAL THROMBECTOMY PROCEDURE AND A MIDDLE CEREBRAL ARTERY (MCA) STENOSIS WAS NOTICED UPON REMOVAL OF THE CLOT. THE PHYSICIAN WANTED TO ANGIOPLASTY THE STENOSIS BEFORE PLACING THE STENT. DURING THE ANGIOPLASTY, THE HYPERGLIDE BALLOON WAS INFLATED IN A VESSEL THAT WAS 1.7MM DISTAL AND 2.5MM PROXIMAL TO ITS NOMINAL SIZE AND THAT WAS WHEN THE REPORTED EVENT OCCURRED WITH THE BALLOON. THE BALLOON WAS, SUBSEQUENTLY, REMOVED FROM THE VESSEL WITH DIFFICULTY. UPON REMOVAL FROM THE BODY, THE BALLOON WAS BROKEN BY THE SCRUB TECHNICIAN TO ENSURE THAT THE WIRE HAD NOT BROKEN. THE BALLOON WAS TESTED PRIOR TO USE, AND IT PERFORMED AS INTENDED DURING TESTING. 10-15MM THE GUIDEWIRE TIP WAS ADVANCED OUT OF THE CATHETER TIP DURING THE INFLATION. THE CONTRAST RATIO WAS 50/50. ONLY ONE TIME THE BALLOON GET INFLATED AND DEFLATED. THE INJECTION RATE WAS STEADY. NON CADENCE, REGULAR 1ML SYRINGE WAS USED. THE PATIENT VESSEL WAS NORMAL.

    Patient problems: No Known Impact Or Consequence To Patient; Paresis

    MAUDE report 8612780 ↗

FDA clearance and classification

Cleared through 510(k) K101570 on Jul 13, 2010 (decision: Substantially Equivalent). Submitted by Micro Therapeutics Dba Ev3 Neurovascular as “Hyperglide And Hyperform Occlusion Balloon Systems”. The FDA review took 39 days. FDA record ↗

Cleared through 510(k) K091458 on Oct 15, 2009 (decision: Substantially Equivalent). Submitted by Ev3 Neurovascular as “Hyperglide Occlusion Balloon System, Hyperform Occlusion Balloon System, Model 104-4112, 104-4113, 104-4127,104-4132,104”. The FDA review took 150 days. FDA record ↗

The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HyperGlideTM 104-4127

Has the HyperGlideTM 104-4127 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00847536005020), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HyperGlideTM 104-4127?

The FDA's MAUDE database lists 1 adverse event report for this device, received on May 15, 2019. That report lists device problem “Burst Container or Vessel” (1); patient problem “Paresis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HyperGlideTM 104-4127?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HyperGlideTM 104-4127 FDA approved?

It was cleared through 510(k) K091458 on Oct 15, 2009; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the HyperGlideTM 104-4127, and is it still sold?

The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand HyperGlideTM, model 104-4127). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, intravascular occluding, temporary devices

All Catheter, intravascular occluding, temporary devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page