PIPELINE PED-300-12Recalls & reported problems
Micro Therapeutics, Inc. · PED-300-12 PIPELINE FLEX 3.00MM X 12MM · Class 3 · Implant
All models of PIPELINE (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00847536014732), received between May 14, 2020 and May 12, 2021 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
11 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jul 13, 2021Open, Classified
Due to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.
Z-2402-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 12, 2021Break; Break
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WAS NO DAMAGE TO THE PHENOM 27 CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11812847 ↗MalfunctionDec 8, 2020Material Deformation; Material Deformation
MEDTRONIC RECEIVED A REPORT THAT THE PIPELINE PUSHWIRE WAS DEFORMED DURING DELIVERY. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN AMORPHOUS, UNRUPTURED ANEURYSM WITH A MAX DIAMETER OF 3MM AND A 2MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MINIMAL. IT WAS UNKNOWN IF DUAL ANTIPLATELET THERAPY (DAPT) WAS ADMINISTERED. IT WAS REPORTED THAT WHEN THE PIPELINE WAS ADVANCED THROUGH THE MICROCATHETER, THE DISTAL TIP WAS VISUALIZED FLUOROSCOPICALLY AND THE PHYSICIAN NOTED THAT THE WIRE APPEARED TO BE DEFORMED. WHEN THE DEVICE WAS REMOVED FROM THE MICROCATHETER, THE DISTAL TIP WAS NOT STRAIGHT. THE PUSHWIRE DID NOT APPEAR TO BE DAMAGED DURING PREPARATION AND DELIVERY; HOWEVER, IT APPEARED TO HAVE BEEN DAMAGED PRIOR TO DELIVERY. NO FRICTION AT ALL WAS NOTED. A REPLACEMENT WAS USED TO COMPLETE THE PROCEDURE. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ANCILLARY DEVICES INCLUDE A PHENOM PLUS, PHENOM 27, SYNCHRO II, TARGET NANO 2 COILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10972013 ↗InjuryMay 14, 2020Break; Break
MEDTRONIC RECEIVED INFORMATION REGARDING A PIPELINE PUSHWIRE THAT BROKE AT THE PROXIMAL MARKER. THE PATIENT WAS BEING TREATED FOR AN UNRUPTURED SACCULAR ANEURYSM OF A POSTERIOR COMMUNICATING ARTERY. THE ANEURYSM MAX DIAMETER WAS 6MM AND THE NECK DIAMETER WAS 3MM. VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT ONE PIPELINE WAS PLACED BUT MISSED THE DISTAL LANDING ZONE SO A SECOND PIPELINE WAS USED. AFTER THE SECOND PIPELINE WAS DEPLOYED, THE SURGEON WAS PULLING OUT THE PUSHWIRE WHEN IT DETACHED AT THE PROXIMAL MARKER AND WAS LEFT IN THE PATIENT. THE SURGEON ATTEMPT TO USE A SNARE TO REMOVE THE PUSHWIRE FRAGMENT FOR AN HOUR BUT COULD NOT REMOVE IT. A STENT WAS PLACED TO TACK DOWN THE WIRE. THE WIRE REMAINS IN THE PATIENT BUT NO FURTHER RELATED SYMPTOMS WERE REPORTED. ANCILLARY DEVICES: PHENOM MICROCATHETER, NAVIEN-058 INTRACRANIAL SUPPORT CATHETER, NON-MEDTRONIC STENT, SHEATH AND GUIDEWIRE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10058289 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-1245-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice