PIPELINE PED-350-18Recalls & reported problems
Micro Therapeutics, Inc. · PED-350-18 PIPELINE FLEX 3.50MM X 18MM · Class 3 · Implant
All models of PIPELINE (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00847536015364), received between Nov 19, 2019 and Apr 21, 2022 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
11 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jul 13, 2021Open, Classified
Due to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.
Z-2402-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure2
- Adverse Event Without Identified Device or Use Problem1
- Migration1
- Packaging Problem1
Patient problems
- Intracranial Hemorrhage1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Obstruction/Occlusion1
- Stroke/CVA1
- Unspecified Nervous System Problem1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2022Packaging Problem; Migration; Packaging Problem; Migration
MEDTRONIC RECEIVED A REPORT THAT THE TIP OF THE PIPELINE OPENED SMOOTHLY. THE DEVICE WAS PULLED BACK TO THE ANCHOR POINT, AND THE MIDDLE AND PROXIMAL OF THE STENT WAS CONTINUED TO BE OPENED SMOOTHLY. AFTER BEING COMPLETELY DEPLOYED, THE ENTIRE STENT SUDDENLY SHIFTED TO THE DISTAL AND MOVED TO THE BIFURCATION OF THE END OF THE INTERNAL CAROTID ARTERY. THE SURGEON SUSPECTED THAT THE STENT WAS NOT THE 350-18 SIZE AS INDICATED ON THE PACKAGE. THE STENT WAS DEPLOYED AND COULD NOT BE REMOVED. THE PIPELINE WAS NOT USED OFF LABEL AND THE PIPELINE WAS PREPARED AS INDICATED IN THE IFU. THERE WERE NO SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THE EVENT. IT WAS UNKNOWN IF A DAPT (DUAL ANTIPLATELET TREATMENT) WAS ADMINISTERED. THE ANGIOGRAPHIC RESULT POST PROCEDURE SHOWED FRONTAL AND LATERAL ANGIOGRAPHY SHOWING SMOOTH BLOOD FLOW. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM IN THE POSTERIOR COMMUNICATING SEGMENT OF THE RIGHT INTERNAL CAROTID ARTERY WITH A MAX DIAMETER OF 3.2MM AND A 2.6MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MODERATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14168368 ↗InjuryMay 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE PATIENT'S CONDITION WAS UNCHANGED. THEY REMAINED ALIVE AND UNCONSCIOUS IN THE ICU.
Patient problems: Obstruction/Occlusion; Stroke/CVA; Unspecified Nervous System Problem
MAUDE report 11806275 ↗InjuryMar 24, 2021Activation Failure; Activation Failure
ADDITIONAL INFORMATION RECEIVED REPORTED THAT SURGICAL INTERVENTION WAS PERFORMED TO RESOLVE THE CEREBRAL HEMORRHAGE, AND THE PATIENT HAS RECOVERED.
Patient problems: Intracranial Hemorrhage
MAUDE report 11562775 ↗MalfunctionNov 19, 2019Activation Failure; Activation Failure
MEDTRONIC RECEIVED REPORT THE DURING THE RELEASE PROCESS, THE TIP OF PIPELINE COULD NOT BE COMPLETELY OPENED. IT WAS RELEASED AGAIN AFTER WITHDRAWN, BUT IT STILL COULD NOT BE OPENED. THE SURGERY WAS COMPLETED SUCCESSFULLY AFTER REPLACING THE PIPELINE WITH A NEW 350-20 PIPELINE. THE DEVICES WERE PREPARED AND USED PER THE INSTRUCTIONS FOR USE (IFU). THIS EVENT OCCURRED DURING THE TREATMENT OF A LEFT INTERNAL CAROTID BED SEGMENT. THE ANEURYSM WAS UNRUPTURED, SACCULAR, THE MAX DIAMETER WAS 3.6MM WITH A NECK DIAMETER OF 2.6MM. THE DISTAL LANDING ZONE WAS 3MM AND THE PROXIMAL WAS 3.5MM. THE VESSEL ANATOMY WAS MINIMAL IN TORTUOSITY. NO PATIENT INJURY OCCURRED. THE PIPELINE WAS NOT PLACED IN A BEND, THE PIPELINE OPENED IN A STRAIGHT SECTION AND WAS PLACED IN A RELATIVELY FLAT, WITHOUT LARGE ARC BENDING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9341866 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-1245-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PE
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page