Skip to content

PipelineTM Vantage Embolization Device with Shield TechnologyTM PED3-027-500-40Recalls & reported problems

Micro Therapeutics, Inc. · STENT PED3-027-500-40 V03 VANTAGE · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536036963), received between Feb 26, 2025 and Feb 26, 2025 — classified as 1 death.

The most frequently reported specific device problem is “Activation Failure” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

11 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 30, 2025Class IOpen, Classified

    Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

    Z-1245-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure1
  • Activation, Positioning or Separation Problem1
  • Collapse1

Patient problems

  • Obstruction/Occlusion1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 26, 2025Collapse; Activation, Positioning or Separation Problem; Activation Failure; Collapse; Activation, Positioning or Separation Problem; Activation Failure

    ADDITIONAL INFORMATION RECEIVED REPORTED THE PHYSICIAN DID NOT FEEL THAT THE COMPLICATION WAS RELATED TO THE PIPELINE VANTAGE DEVICE. INSTEAD, THE PHYSICIAN FELT LIKE THE PATIENT WAS ABLE TO LIVE FOR ADDITIONAL TIME BECAUSE OF THE PROCEDURE THAT WAS PERFORMED. THE PATIENT HAD UNDERGONE TREATMENT OF A FUSIFORM DISSECTING GIANT VERTEBRAL ARTERY ANEURYSM THAT WAS VERY CHALLENGING TO TREAT WHICH WAS KNOWN OFF-LABEL USE OF THE PIPELINE VANTAGE. THE IMPLANTATION OF THE PIPELINE WENT EVEN BETTER THAN EXPECTED, WITH ONLY MINOR CHALLENGES. THE PATIENT'S CONDITION IMPROVED OVER TIME POST-OPERATIVELY, HOWEVER, AT 6 MONTH FOLLOW UP THE PATIENT REPORTED SYMPTOMS AND UPON ANGIOGRAPHIC IMAGING, THROMBOSIS AROUND THE IMPLANT AND NARROWING OF THE DEVICES WAS IDENTIFIED. THE PATIENT EXPIRED AROUND 9 MONTHS POST-IMPLANT. THE SOURCE OF THE IN-STENT STENOSIS WAS UNCLEAR BUT THE PHYSICIAN DID NOT ATTRIBUTE IT TO THE PIPELINE AS IT WAS NOTED THE PATIENT HAD OTHER CONDITIONS, AND THE PHYSICIAN CONSIDERED IT MORE LIKELY THAT THE PIPELINE DEVICE CAPTURED SOME THROMBUS, RATHER THAN GENERATING IT. THE PHYSICIAN EMPHASIZED MULTIPLE TIMES THAT THE PATIENT DEATH WAS, ON THEIR PART, NOT CONSIDERED TO BE RELATED T

    Patient problems: Obstruction/Occlusion

    MAUDE report 21472459 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.

  • Z-1246-2025 · Jan 30, 2025 · Open, Classified
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
  • Z-2403-2021 · Jul 13, 2021 · Open, Classified
    EV3 Pipeline Flex Embolization Device with Flex Shield Technology
  • Z-2402-2021 · Jul 13, 2021 · Open, Classified
    Pipeline Flex Embolization Device
  • Z-1527-2020 · Feb 14, 2020 · Terminated
    ev3 Pipeline Flex Embolization Device. For neurological endovascular use.
  • Z-0310-2017 · Oct 5, 2016 · Terminated
    X-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
  • Z-0308-2017 · Oct 5, 2016 · Terminated
    Alligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
  • Z-0307-2017 · Oct 5, 2016 · Terminated
    Pipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
  • Z-0309-2017 · Oct 5, 2016 · Terminated
    Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
  • Z-0120-2016 · Sep 4, 2015 · Terminated
    -Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
  • Z-1453-2014 · Apr 1, 2014 · Terminated
    Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
  • Z-2045-2013 · Jul 19, 2013 · Terminated
    Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck

Other Intracranial aneurysm flow diverter devices

All Intracranial aneurysm flow diverter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page