PipelineTM Vantage Embolization Device with Shield TechnologyTM PED3-027-550-16Recalls & reported problems
Micro Therapeutics, Inc. · STENT PED3-027-550-16 V03 VANTAGE · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536037007), received between Oct 2, 2024 and Dec 10, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Activation Failure” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
11 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jan 30, 2025Class IOpen, Classified
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
Z-1245-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure1
- Collapse1
- Migration1
Patient problems
- Intracranial Hemorrhage1
- Ischemia Stroke1
- Muscle Weakness/Atrophy1
- No Clinical Signs, Symptoms or Conditions1
- Paresis1
- Thrombosis/Thrombus1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2024Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A NEUROVASCULAR FLOW DIVERSION PROCEDURE, AN UNRUPTURED SACCULAR ANEURYSM LOCATED IN THE OPHTHALMIC INTERNAL CAROTID ARTERY WITH A MAXIMUM DIAMETER OF 5.1 MILLIMETERS AND A NECK DIAMETER OF 4.0 MILLIMETERS WAS BEING TREATED. THE PATIENT HAD SEVERE VESSEL TORTUOSITY. THE PIPELINE EMBOLIZATION DEVICE (PED2-475-12) FAILED TO OPEN IN THE DISTAL SECTION AND BECAME STUCK IN THE PHENOM 27 MICROCATHETER UPON REMOVAL. THEY HAD ATTEMPTED TO REPOSITION THE PIPELINE WITH THE MICROCATHETER IN AN ATTEMPT TO OPEN IT. MORE THAN 50% OF THIS DEVICE WAS DEPLOYED BEFORE THE FAILURE, AND IT WAS RESHEATHED LESS THAN OR EQUAL TO TWO TIMES BEFORE BEING REMOVED FROM THE PATIENT. THE DEVICE HAD NOT BEEN POSITIONED IN A BEND. ADDITIONALLY, A PIPELINE VANTAGE WITH SHIELD TECHNOLOGY (PED3-027-550-16) FAILED TO OPEN IN THE MIDDLE SECTION. THE DISTAL SEGMENT HAD BEEN POSITIONED IN A BEND. MORE THAN 50% OF THIS DEVICE WAS DEPLOYED, AND IT WAS RESHEATHED MORE THAN TWO TIMES BEFORE REMOVAL. ANOTHER PIPELINE VANTAGE 5X14 WAS SUCCESSFULLY IMPLANTED IN THE PATIENT. THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IF
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20893491 ↗InjuryOct 2, 2024Collapse; Migration; Collapse; Migration
ADDITIONAL INFORMATION RECEIVED REPORTED AFTER IMPLANTATION OF THE PIPELINE THREE RUNS OF CONTRAST WERE DONE TO CONFIRM PIPELINE LOCATION AND APPOSITION. ALL OF THESE SHOWED GOOD APPOSITION AND PROPER LOCATION OF THE DEVICE. 10 MINUTES AFTER THE PROCEDURE, THE PATIENT AWOKE AND EXPERIENCED INABILITY TO MOVE RIGHT SIDE EXTREMITIES. CTA SHOWED DISPLACEMENT AND MIGRATION OF THE PIPELINE VANTAGE INTO THE RIGHT M1 ICA SEGMENT. IMAGING REVEALED THROMBUS FORMATION IN THE LUMEN OF THE PIPELINE DEVICE. A BOLUS OF INTEGRILIN AND A SUBSEQUENT INFUSION OF THIS MEDICATION WAS INITIATED. A REPEAT PHYSICAL EXAM REVEALED AN IMPROVEMENT OF THE LEFT HEMIPARESIS. A REPEAT ANGIOGRAM SHOWED A DISTAL MIGRATION OF THE PIPELINE TOWARDS THE DISTAL CAVERNOUS SEGMENT, BOTH PORTIONS OF THE DEVICE, DISTAL AND PROXIMAL MIGRATED. THE PROXIMAL END OF THE DEVICE COVERAGE OF THE NECK OF THE ANEURYSM WAS MINIMAL AND AROUND 1-3MM. A SMALL RESIDUAL THROMBUS WAS OBSERVED WITHIN THE DEVICE AND THERE WAS GOOD FLOW IN THE CAROTID. FOR THESE REASONS, IT WAS DECIDED THAT IT WAS BEST TO LEAVE THE DEVICE ALONE. IT WAS THOUGHT THAT ANY SMALL THROMBUS AROUND THE DEVICE WOULD LYSE DUE TO THE PATIENT'S ANTIPLATELET REGIMEN. AT AR
Patient problems: Intracranial Hemorrhage; Ischemia Stroke; Muscle Weakness/Atrophy; Paresis; Thrombosis/Thrombus; Vascular Dissection
MAUDE report 20352069 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.
- Z-1246-2025 · Jan 30, 2025 · Open, ClassifiedMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
- Z-2402-2021 · Jul 13, 2021 · Open, ClassifiedPipeline Flex Embolization Device
- Z-2403-2021 · Jul 13, 2021 · Open, ClassifiedEV3 Pipeline Flex Embolization Device with Flex Shield Technology
- Z-1527-2020 · Feb 14, 2020 · Terminatedev3 Pipeline Flex Embolization Device. For neurological endovascular use.
- Z-0309-2017 · Oct 5, 2016 · TerminatedMarathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
- Z-0310-2017 · Oct 5, 2016 · TerminatedX-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
- Z-0307-2017 · Oct 5, 2016 · TerminatedPipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
- Z-0308-2017 · Oct 5, 2016 · TerminatedAlligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
- Z-0120-2016 · Sep 4, 2015 · Terminated-Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
- Z-1453-2014 · Apr 1, 2014 · TerminatedPipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
- Z-2045-2013 · Jul 19, 2013 · TerminatedPipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page