PipelineTM Vantage Embolization Device with Shield TechnologyTM PED3-027-550-16Recalls & reported problems

Micro Therapeutics, Inc. · STENT PED3-027-550-16 V03 VANTAGE · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536037007), received between Oct 2, 2024 and Dec 10, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

11 other recalls cover devices cleared under the same FDA submission (P100018); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 30, 2025Class IOpen, Classified

    Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

    Z-1245-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure1
  • Collapse1
  • Migration1

Patient problems

  • Intracranial Hemorrhage1
  • Ischemia Stroke1
  • Muscle Weakness/Atrophy1
  • No Clinical Signs, Symptoms or Conditions1
  • Paresis1
  • Thrombosis/Thrombus1
  • Vascular Dissection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 10, 2024Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A NEUROVASCULAR FLOW DIVERSION PROCEDURE, AN UNRUPTURED SACCULAR ANEURYSM LOCATED IN THE OPHTHALMIC INTERNAL CAROTID ARTERY WITH A MAXIMUM DIAMETER OF 5.1 MILLIMETERS AND A NECK DIAMETER OF 4.0 MILLIMETERS WAS BEING TREATED. THE PATIENT HAD SEVERE VESSEL TORTUOSITY. THE PIPELINE EMBOLIZATION DEVICE (PED2-475-12) FAILED TO OPEN IN THE DISTAL SECTION AND BECAME STUCK IN THE PHENOM 27 MICROCATHETER UPON REMOVAL. THEY HAD ATTEMPTED TO REPOSITION THE PIPELINE WITH THE MICROCATHETER IN AN ATTEMPT TO OPEN IT. MORE THAN 50% OF THIS DEVICE WAS DEPLOYED BEFORE THE FAILURE, AND IT WAS RESHEATHED LESS THAN OR EQUAL TO TWO TIMES BEFORE BEING REMOVED FROM THE PATIENT. THE DEVICE HAD NOT BEEN POSITIONED IN A BEND. ADDITIONALLY, A PIPELINE VANTAGE WITH SHIELD TECHNOLOGY (PED3-027-550-16) FAILED TO OPEN IN THE MIDDLE SECTION. THE DISTAL SEGMENT HAD BEEN POSITIONED IN A BEND. MORE THAN 50% OF THIS DEVICE WAS DEPLOYED, AND IT WAS RESHEATHED MORE THAN TWO TIMES BEFORE REMOVAL. ANOTHER PIPELINE VANTAGE 5X14 WAS SUCCESSFULLY IMPLANTED IN THE PATIENT. THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IF

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20893491 ↗
  • InjuryOct 2, 2024Collapse; Migration; Collapse; Migration

    ADDITIONAL INFORMATION RECEIVED REPORTED AFTER IMPLANTATION OF THE PIPELINE THREE RUNS OF CONTRAST WERE DONE TO CONFIRM PIPELINE LOCATION AND APPOSITION. ALL OF THESE SHOWED GOOD APPOSITION AND PROPER LOCATION OF THE DEVICE. 10 MINUTES AFTER THE PROCEDURE, THE PATIENT AWOKE AND EXPERIENCED INABILITY TO MOVE RIGHT SIDE EXTREMITIES. CTA SHOWED DISPLACEMENT AND MIGRATION OF THE PIPELINE VANTAGE INTO THE RIGHT M1 ICA SEGMENT. IMAGING REVEALED THROMBUS FORMATION IN THE LUMEN OF THE PIPELINE DEVICE. A BOLUS OF INTEGRILIN AND A SUBSEQUENT INFUSION OF THIS MEDICATION WAS INITIATED. A REPEAT PHYSICAL EXAM REVEALED AN IMPROVEMENT OF THE LEFT HEMIPARESIS. A REPEAT ANGIOGRAM SHOWED A DISTAL MIGRATION OF THE PIPELINE TOWARDS THE DISTAL CAVERNOUS SEGMENT, BOTH PORTIONS OF THE DEVICE, DISTAL AND PROXIMAL MIGRATED. THE PROXIMAL END OF THE DEVICE COVERAGE OF THE NECK OF THE ANEURYSM WAS MINIMAL AND AROUND 1-3MM. A SMALL RESIDUAL THROMBUS WAS OBSERVED WITHIN THE DEVICE AND THERE WAS GOOD FLOW IN THE CAROTID. FOR THESE REASONS, IT WAS DECIDED THAT IT WAS BEST TO LEAVE THE DEVICE ALONE. IT WAS THOUGHT THAT ANY SMALL THROMBUS AROUND THE DEVICE WOULD LYSE DUE TO THE PATIENT'S ANTIPLATELET REGIMEN. AT AR

    Patient problems: Intracranial Hemorrhage; Ischemia Stroke; Muscle Weakness/Atrophy; Paresis; Thrombosis/Thrombus; Vascular Dissection

    MAUDE report 20352069 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100018. One submission often covers many models, so check the product and lot details in each record.

  • Z-1246-2025 · Jan 30, 2025 · Open, Classified
    Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PE
  • Z-2402-2021 · Jul 13, 2021 · Open, Classified
    Pipeline Flex Embolization Device
  • Z-2403-2021 · Jul 13, 2021 · Open, Classified
    EV3 Pipeline Flex Embolization Device with Flex Shield Technology
  • Z-1527-2020 · Feb 14, 2020 · Terminated
    ev3 Pipeline Flex Embolization Device. For neurological endovascular use.
  • Z-0309-2017 · Oct 5, 2016 · Terminated
    Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled
  • Z-0310-2017 · Oct 5, 2016 · Terminated
    X-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 10
  • Z-0307-2017 · Oct 5, 2016 · Terminated
    Pipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracrani
  • Z-0308-2017 · Oct 5, 2016 · Terminated
    Alligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04
  • Z-0120-2016 · Sep 4, 2015 · Terminated
    -Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based
  • Z-1453-2014 · Apr 1, 2014 · Terminated
    Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
  • Z-2045-2013 · Jul 19, 2013 · Terminated
    Pipeline Embolization Device (PED) The Pipeline Embolization Device is indicated for the endovascular treatment of adults (22 years of age and older) with large or giant wide-neck