PhenomTM 21 Catheter FG13150-0615-2S (UDI 00847536041813)Recalls & reported problems
Micro Therapeutics, Inc. · CATHETER FG13150-0615-2S V04 · Class 2
All models of PhenomTM 21 Catheter (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536041813), received between Aug 31, 2021 and Aug 31, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Deformation Due to Compressive Stress1
- Physical Resistance/Sticking1
Patient problems
- Bradycardia1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 31, 2021Deformation Due to Compressive Stress; Physical Resistance/Sticking; Deformation Due to Compressive Stress; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE DELIVERY SYSTEM AND PIPELINE VANTAGE WERE PREPARED PER THE INSTRUCTIONS FOR USE (IFU) AND DELIVERED TO THE TARGET LOCATION SMOOTHLY WITH NO ISSUES. THE STENT WAS DEPLOYED WITH NO ISSUES, AND THE STENT OPENED WELL. THE PROXIMAL END REQUIRED SLIGHT WAGGING TO FULLY APPOSE. THE PHYSICIAN DID NOT WANT TO TAKE THE CATHETER FULLY THROUGH TO CAPTURE THE SLEEVES, AND THE RHV HAD NOT BEEN LOCKED DOWN. WHEN INJECTING THE CONTRAST FOR A RUN, THE MICROCATHETER MOVED FORWARD THROUGH THE CENTER OF THE BRAID INDEPENDENTLY, AND WAS BEYOND THE DISTAL END OF THE BRAID. WHEN TRYING TO PULL THE DELIVERY WIRE AND SLEEVES TO RECAPTURE, IT WAS NOTED TO BE DIFFICULT AND THERE WAS RESISTANCE. THE CATHETER AND WIRE WERE SUCCESSFULLY REMOVED CAREFULLY. ONCE REMOVED, IT WAS NOTICED THE PHENOM HAD ACCORDIONED, AND IT WAS DIFFICULT TO PULL THE WIRE BACK THROUGH. IT WAS NOTED THE PATIENT HAD BRADYCARDIA DURING DEPLOYMENT OF TWO OCCASIONS WHICH RESOLVED ON DRAWING BACK THE CAT 5 NO FURTHER ISSUES. THERE WAS NO OTHER INJURY TO THE PATIENT AS A RESULT OF THE EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN AMORPHOUS, UNRUPTURED ANEURYSM OF THE ANTERIOR COMMUNICATING
Patient problems: Bradycardia
MAUDE report 12393912 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page