PhenomTM 27 Catheter FG15150-0630-1S (UDI 00847536041875)Recalls & reported problems
Micro Therapeutics, Inc. · CATHETER FG15150-0630-1S V04 · Class 2
All models of PhenomTM 27 Catheter (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00847536041875), received between Dec 3, 2019 and Dec 20, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Physical Resistance/Sticking2
- Adverse Event Without Identified Device or Use Problem1
- No Apparent Adverse Event1
Patient problems
- Fistula1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 20, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
2022-OCT-24 MPXR 996323 (REP, HCP, FOR, DIS): MEDTRONIC RECEIVED REPORT THAT THE PIPELINE JAMMED IN THE HUB OF THE PHENOM 27 MICROCATHETER. IT WAS NOTED THAT THE DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE TIP OF THE SHEATH WAS SEATED SECURELY IN THE CATHETER HUB. THE CATHETER WAS FLUSHED PER IFU. BOTH DEVICES WERE REPLACED WITH ANOTHER MANUFACTURER'S PRODUCTS TO COMPLETE THE PROCEDURE. THERE WAS NO NOTED PATIENT INVOLVEMENT. 2022-11-08 E1 (FOR, REP, HCP): NEW INFORMATION WAS RECEIVED. THERE WAS NO DAMAGE OBSERVED ON THE PIPELINE OR CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16015412 ↗MalfunctionNov 16, 2021No Apparent Adverse Event; No Apparent Adverse Event
MEDTRONIC RECEIVED A REPORT THAT THE PIPELINE FAILED TO OPEN IN THE MIDDLE SECTION AND WAS DAMAGED. THE PATIENT WAS UNDERGOING TREATMENT FOR AN UNRUPTURED, SACCULAR ANEURYSM LOCATED IN THE RIGHT OPHTHALMIC ARTERY SEGMENT. THE MAX DIAMETER WAS 7MM, AND THE NECK DIAMETER WAS 6MM. THE PATIENT'S VESSEL TORTUOSITY WAS SEVERE. THE LANDING ZONE WAS 4MM DISTAL AND 4.5MM PROXIMAL. THE ACCESS VESSEL WAS THE FEMORAL ARTERY. IT WAS REPORTED THAT AFTER THE MICROCATHETER WAS IN PLACE AND THE DOCTOR STARTED TO DELIVER THE PIPELINE IT WAS FOUND THAT THE STENT WAS KINKED AT THE HORIZONTAL SECTION OF C4, AND THE ANTERIOR AND POSTERIOR SITES COULD BE OPENED. THE KINK AND FLATTENED HORIZONTAL SECTION OF THE STENT COULD NOT BE RESOLVED BY PUSHING AND PULLING THE STENT SEVERAL TIMES. THE PIPELINE WAS REPEATEDLY RETRIEVED AND REPOSITIONED TO THE DISTAL SIDE. AFTER MORE THAN TWO RESHEATHING ATTEMPTS, IT WAS STILL KINKED AT THE C4 HORIZONTAL SEGMENT AND COULD NOT BE DEPLOYED NORMALLY. IT WAS NOTED THAT THERE WERE BURRS ON THE TIP OF THE PIPELINE, SO THE ENTIRE PIPELINE AND MICROCATHETER DELIVERY SYSTEM WERE WITHDRAWN. THE PIPELINE WAS NOT POSITIONED IN A BEND WHEN THE MIDDLE SECTION FAILED TO OPEN, AND MOR
Patient problems: Insufficient Information
MAUDE report 12820146 ↗MalfunctionNov 4, 2020Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE PIPELINE EXPERIENCED RESISTANCE AND TWISTING DURING DELIVERY. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN AMORPHOUS, UNRUPTURED ANEURYSM OF THE SUPERCLINOID ICA. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS SEVERE. IT WAS REPORTED THAT THE DEVICES WERE PREPARED PER THE INSTRUCTIONS FOR USE (IFU). SEVERE TORTUOSITY WAS MET WHILE ADVANCING THE PIPELINE, AND DIFFICULTY WAS MET. THE PIPELINE TWISTED DISTALLY DESPITE RE-SHEATHING, AND ALSO TWISTED PROXIMALLY DESPITE RE-SHEATHING DISTAL TO TWIST. THE PHYSICIAN ATTEMPTED TO RE-POSITION AND REDEPLOY THE PIPELINE, BUT TWISTING CONTINUED TO OCCUR. THE PIPELINE AND PHENOM MICROCATHETER WERE THEN REMOVED. A NEW MICROCATHETER AND DEVICE WERE THEN USED WITH SUCCESS. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THE EVENT. POST-PROCEDURE ANGIOGRAPHIC RESULTS SHOWED A SUCCESSFUL PROCEDURE. ANCILLARY DEVICES INCLUDE A NEURON MAX, PHENOM PLUS, PHENOM 27, FATHOM 16. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE RESISTANCE WAS THROUGHOUT THE CATHETER, AND NO DAMAGE WAS OBSERVED TO THE PUSHWIRE OR CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10786271 ↗InjuryDec 3, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED A REPORT THAT A PATIENT DEVELOPED A FISTULA DURING A PROCEDURE. THE PATIENT WAS UNDERGOING EMBOLIZATION TREATMENT FOR AN UNRUPTURED AMORPHOUS ANEURYSM LOCATED IN RIGHT INTERNAL CAROTID ARTERY (ICA) LANDING ZONE DISTAL 3.2MM PROXIMAL 4.7MM. THE VESSEL WAS OBSERVED MODERATELY TORTUOUS. IT WAS REPORTED THAT THE DEVICES WERE PREPARED AS INDICATED IN THE IFU. THE PHYSICIAN ADVANCED A GUIDE CATHETER AND MICROCATHETER. A RUN WITH CONTRAST REVEALED A FISTULA THAT THE PHYSICIAN BELIEVES WAS CATHETER INDUCED. THE PHYSICIAN DECIDED TO ¿SACK¿ THE VESSEL AS THE RIGHT ICA ON IMAGING WAS BEING WELL FED FROM CONTRALATERAL FLOW. RIGHT ICA WAS SACRIFICED; THE PATIENT¿S CONTRALATERAL FLOW WAS FEEDING RIGHT CIRCULATION. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT.
Patient problems: Fistula
MAUDE report 9407275 ↗
FDA clearance and classification
Cleared through 510(k) K151638 on Nov 13, 2015 (decision: Substantially Equivalent). Submitted by Cathera, Inc. as “Phenom 17 Catheter, Phenom 21 Catheter, Phenom 27 Catheter, Phenom Plus Catheter”. The FDA review took 149 days. FDA record ↗
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PhenomTM 27 Catheter FG15150-0630-1S
Has the PhenomTM 27 Catheter FG15150-0630-1S been recalled?
No FDA recall record in our data names this exact device identifier (DI 00847536041875), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PhenomTM 27 Catheter FG15150-0630-1S?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Dec 3, 2019 and Dec 20, 2022. In those reports, the most common specific device problem is “Physical Resistance/Sticking” (2); the most common patient problem is “Fistula” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PhenomTM 27 Catheter FG15150-0630-1S?
None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PhenomTM 27 Catheter FG15150-0630-1S FDA approved?
It was cleared through 510(k) K151638 on Nov 13, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the PhenomTM 27 Catheter FG15150-0630-1S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the PhenomTM 27 Catheter FG15150-0630-1S, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand PhenomTM 27 Catheter, model FG15150-0630-1S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page