VARIANT nbs 250-3000Recalls & reported problems
Bio-Rad Laboratories, Inc. · VARIANT nbs Sickle Cell Program Reorder Pack · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Apr 7, 2022Class IIOpen, Classified
The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CDROM parameters for the intended use. However, there is a risk that the customer's custom pattern settings may be different than the default settings, resulting in a different pattern being assigned to some samples that the laboratory has validated. Following the complaints from 4 customers who experienced these issues, the firm initiated this recall.
Z-1208-2022 · Root cause: Packaging process control · Recall details · FDA record ↗
FDA clearance and classification
Cleared through 510(k) K051072 on May 12, 2005 (decision: Substantially Equivalent). Submitted by Bio-Rad Laboratories, Inc. as “Bio-Rad Variant Nbs Sickle Cell Program”. The FDA review took 16 days. FDA record ↗
The FDA regulates product code GKA (abnormal hemoglobin quantitation) under 21 CFR 864.7415 as a Class II device. 21 CFR 864.7415 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VARIANT nbs 250-3000
Has the VARIANT nbs 250-3000 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1208-2022 (Class II), began on Apr 7, 2022: The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the VARIANT nbs 250-3000 FDA approved?
It was cleared through 510(k) K051072 on May 12, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the VARIANT nbs 250-3000, and is it still sold?
The FDA's device database lists Bio-Rad Laboratories, Inc. as the company responsible for it (brand VARIANT nbs, model 250-3000). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page