Skip to content

Autoimmune EIA 425-2260Recalls & reported problems

Bio-Rad Laboratories, Inc. · Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Apr 3, 2019Completed

    An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged.

    Z-1800-2020 · Root cause: Under Investigation by firm · FDA record ↗

Sources and updates