Skip to content

GS 25256Recalls & reported problems

Bio-RAD Laboratories, Inc. · Human Immunodeficiency Virus Types 1 and 2 (Recombinant and Synthetic Peptide) GS HIV-1/HIV-2 PLUS O EIA 4800 Test Kit · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847865001953), received between Oct 14, 2019 and Oct 14, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method1

Patient problems

  • Test Result1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 14, 2019Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ON SEPTEMBER 19, 2019 BIO-RAD TECHNICAL SUPPORT RECEIVED A COMPLAINT FORWARDED FROM A DISTRIBUTION PARTNER ABOUT AN INCIDENT THAT HAPPENED AT CUSTOMER FACILITY. THE INITIAL COMPLAINT WAS FOR AN INCREASE IN INITIALLY REACTIVE (IR) RESULTS THAT DID NOT REPEAT IN SECONDARY TESTING USING THE BIO-RAD GS HIV-1/HIV-2 PLUS O ASSAY. AFTER A PRELIMINARY INVESTIGATION IT WAS FOUND THAT THE IR RESULTS IN QUESTION ALL CAME FROM ONE PLATE RUN ON (B)(6) 2019. FURTHER INVESTIGATION INTO THE ISSUE REVEALED THAT THE RESULTS IN QUESTION WERE DONOR SAMPLES THAT WERE, DUE TO USER ERROR, INADVERTENTLY RUN IN NULL STRIPS IN ROWS 9-12 ON THE TEST PLATE. THE CUSTOMER'S PRACTICE IS TO USE PREVIOUSLY TESTED STRIPS THAT HAVE BEEN CLEANED AS NULL STRIPS. PREVIOUSLY USED STRIPS ARE MARKED WITH A BLACK LINE ACROSS THE BOTTOM OF THE STRIP TO INDICATE THEM AS NULL. THE PACKAGE INSERT STATES THAT THE OPERATOR SHOULD "TAKE CARE WHEN ASSEMBLING PARTIAL PLATES WITH COATED AND UNCOATED (NULL) STRIPS, AS AUTOMATED SYSTEMS CANNOT DISTINGUISH BETWEEN THE STRIPS AND WILL REPORT RESULTS FOR ALL WELLS THAT ARE ASSIGNED A SAMPLE ID NUMBER (EVEN IF A NULL STRIP IS INADVERTENTLY PLACED WHERE SAMPLE IDS HAVE BEEN ASSIGNED)." IT

    Patient problems: Test Result

    MAUDE report 9187846 ↗

Other Test, Hiv Detection devices

All Test, Hiv Detection devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page