GS 26217Recalls & reported problems
Bio-RAD Laboratories, Inc. · GS HIV Combo Ag/Ab EIA, 192 Tests · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847865010658), received between Aug 18, 2016 and May 10, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “False Positive Result” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- False Positive Result1
Patient problems
- Emotional Changes1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 10, 2018False Positive Result
BIO-RAD TECHNICAL SUPPORT WAS CONTACTED BY (B)(6) HOSPITAL ON 4/12/2018, TO REPORT THAT ON (B)(6) 2018 A PATIENT SAMPLE WAS TESTED WITH GS HIV COMBO AG/AB EIA (LOT 159MJJ-02, EXP. 06/08/2018), WHICH GAVE AN INITIAL (B)(6). THE SAMPLE WAS THEN REPEATED ON (B)(6) 2018 IN DUPLICATE WITH GS HIV COMBO AG/AB EIA (LOT 159MJJ-02, EXP. 06/08/2018); BOTH RESULTS WERE (B)(6). THE INITIAL SAMPLE WAS ALSO TESTED ON (B)(6) 2018 ON THE GEENIUS WHICH SHOWED (B)(6) PRESENT, (B)(6) AS THE FINAL INTERPRETATION. THE CUSTOMER REPORTED THAT THE PATIENT WAS ADMITTED TO THE PSYCHIATRIC UNIT DUE TO HER REACTION AFTER RECEIVING THE INITIAL (B)(6) TEST RESULTS. THE CUSTOMER ALSO REPORTED THAT THE PATIENT IS/HAS BEEN A PSYCHIATRIC PATIENT. THE PATIENT REQUESTED A RETEST, AND WAS REDRAWN ON (B)(6) 2018. THE RESULTS FOR THE SECOND SAMPLE FOR GS HIV COMBO AG/AB EIA (LOT 101NJJ-02, EXP. 01/12/2018) GAVE (B)(6) RESULTS. THE CUSTOMER HAD A THIRD SAMPLE COLLECTED FROM THE PATIENT ON (B)(6) 2018, WITH A PSEUDO CHAIN OF CUSTODY. THE RESULTS FOR THE THIRD SAMPLE FOR GS HIV COMBO AG/AB EIA (LOT 101NJJ-02, EXP. 01/12/2018) GAVE (B)(6) RESULTS. NAT TESTING WAS PERFORMED ON THE SECOND AND THIRD SAMPLES COLLECTED FROM THE P
Patient problems: Emotional Changes
MAUDE report 7504774 ↗Aug 18, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2016, THE CUSTOMER CONTACTED BIO-RAD LABORATORIES TECHNICAL SERVICES AND IT WAS DISCOVERED THAT THEY ARE NOT FOLLOWING THE PACKAGE INSERT FOR (B)(6) WITH REGARD TO APPROVED SAMPLE TYPE. THE CUSTOMER IS RUNNING WHOLE BLOOD SAMPLES ON THE ASSAY, WHERE THE ASSAY DOES NOT HAVE A WHOLE BLOOD CLAIM. THE ASSAY IS APPROVED ONLY FOR SERUM OR PLASMA SAMPLES. THERE IS ALSO CONCERN THAT THE CUSTOMER DOES NOT APPEAR TO BE FOLLOWING STANDARD CLSI GP44-A4 SAMPLE HANDLING REQUIREMENTS TO SEPARATE THE SAMPLE FROM THE CLOT AS SOON AS POSSIBLE IN ORDER TO PRESERVE MAXIMUM P24 ANTIGEN REACTIVITY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5886547 ↗
FDA clearance and classification
The device record lists FDA submission BP100064; its details are not in the openFDA clearance files we use.
The FDA regulates product code MZF (test, hiv detection) under 21 CFR 866.3956 as a Class II device. 21 CFR 866.3956 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the GS 26217
Has the GS 26217 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00847865010658), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the GS 26217?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Aug 18, 2016 and May 10, 2018. In those reports, the most common specific device problem is “False Positive Result” (1); the most common patient problem is “Emotional Changes” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GS 26217?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the GS 26217, and is it still sold?
The FDA's device database lists Bio-RAD Laboratories, Inc. as the company responsible for it (brand GS, model 26217). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Test, Hiv Detection devices
- Determine™ HIV–1/2 Ag/Ab Combo 7D2648Alere Scarborough, Inc. · Test, Hiv Detection405 reports
- Alere Determine™ HIV–1/2 Ag/Ab Combo 7D2648Alere Scarborough, Inc. · Test, Hiv Detection238 reports
- Elecsys HIV Duo 08836973190Roche Diagnostics GmbH · Test, HIV detection103 reports
- ARCHITECT 2P36-01Abbott GmbH · Test, hiv detection1 recall0 reports
- Alere Determine™ HIV–1/2 Ag/Ab Combo 7D2649Alere Scarborough, Inc. · Test, Hiv Detection27 reports
- cobas® HIV-1 09040803190Roche Molecular Systems, Inc. · Test, HIV detection24 reports
- VITROS 6842781Ortho Clinical Diagnostics · Test, hiv detection16 reports
- RealTime 6L18-90Abbott Molecular Inc. · Test, hiv detection12 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page