Model EMV+ Vent, Army, MRI 799-EMVP-04Recalls & reported problems

Zoll Medical Corporation · Ventilator · Class 2

All models of Model EMV+ Vent, Army, MRI (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll EMV+ is a ventilator. Of the 606 reports naming it, most describe error messages on the display, failures to power up or self-test, and calibration or mechanical problems found during testing, with no patient involved. The reports include no injuries and 1 death. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
605
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 605 adverse event reports that list this exact device (DI 00847946020620), received between Mar 10, 2017 and Nov 24, 2025 — classified as 1 death, 604 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (150 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Displays Incorrect Message150
  • Failure of Device to Self-Test77
  • Mechanical Problem77
  • Failure to Power Up57
  • Calibration Problem52
  • Device Operational Issue42
  • No Display/Image39
  • Use of Device Problem39

Patient problems

  • No Patient Involvement385
  • No Information122
  • No Clinical Signs, Symptoms or Conditions75
  • No Consequences Or Impact To Patient17
  • Insufficient Information4
  • No Known Impact Or Consequence To Patient3
  • Death1
  • Osteopenia/ Osteoporosis1

Recent adverse event reports

The 25 most recent of 605 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Failure to Power Up; Smoking; Failure to Power Up; Smoking

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP AND SMOKED. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23633958 ↗
  • MalfunctionSep 24, 2025Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23133656 ↗
  • MalfunctionMay 2, 2025Communication or Transmission Problem; Communication or Transmission Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "INTERNAL COMM FAILURE- 1175" ERROR MESSAGE COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21950357 ↗
  • MalfunctionNov 5, 2024No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S DISPLAY WAS MISSING CLINICAL INFORMATION. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20609894 ↗
  • MalfunctionAug 15, 2024Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19991804 ↗
  • MalfunctionMay 14, 2024No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE'S DISPLAY BLANKED OUT. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19308929 ↗
  • MalfunctionMay 8, 2024Communication or Transmission Problem; Communication or Transmission Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "INTERNAL COMM FAILURE -1175" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19273715 ↗
  • MalfunctionApr 4, 2024Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19041023 ↗
  • MalfunctionFeb 26, 2024No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18776505 ↗
  • MalfunctionJan 29, 2024Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18599227 ↗
  • MalfunctionJan 10, 2024Calibration Problem; Calibration Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "RUNTIME CALIBRATION FAILURE - 1051" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18483562 ↗
  • MalfunctionNov 27, 2023Unexpected Shutdown; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A 62-YEAR-OLD MALE PATIENT, THE DEVICE INAPPROPRIATELY SHUT DOWN. COMPLAINANT INDICATED THAT THE CLINICIAN BAG VENTILATED TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18208820 ↗
  • MalfunctionSep 18, 2023Mechanical Problem; Calibration Problem; Mechanical Problem; Calibration Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED "COMPRESSOR FAILURE -1001"AND "RUNTIME CALIBRATION FAILURE - 1051" ERROR MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17763994 ↗
  • MalfunctionJun 30, 2023Communication or Transmission Problem; Communication or Transmission Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED AN "INTERNAL COMM FAILURE 1474" ERROR MESSAGE". COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17237458 ↗
  • MalfunctionJun 27, 2023Calibration Problem; Calibration Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "RUNTIME CALIBRATION FAILURE - 1051" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17209835 ↗
  • MalfunctionJun 27, 2023Calibration Problem; Calibration Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "RUNTIME CALIBRATION FAILURE - 1051" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17209662 ↗
  • MalfunctionMay 19, 2023Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "SELF CHECK FAILURE -1003" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16968976 ↗
  • MalfunctionMay 19, 2023Failure of Device to Self-Test; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "SELF CHECK FAILURE - 1003" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16962038 ↗
  • MalfunctionMay 12, 2023Failure to Power Up; Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16918546 ↗
  • MalfunctionApr 14, 2023Improper Flow or Infusion; Improper Flow or Infusion

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "AIRWAY PRESSURE SENSING FAILURE - 1052" MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16744443 ↗
  • MalfunctionApr 10, 2023Communication or Transmission Problem; Communication or Transmission Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "INTERNAL COMM FAILURE 1474" ERROR MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16707373 ↗
  • MalfunctionMar 28, 2023No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16628363 ↗
  • MalfunctionMar 2, 2023Calibration Problem; Failure of Device to Self-Test; Calibration Problem; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16471494 ↗
  • MalfunctionFeb 22, 2023No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16419196 ↗
  • MalfunctionJan 18, 2023Mechanical Problem; Communication or Transmission Problem; Failure of Device to Self-Test; Mechanical Problem; Communication or Transmission Problem; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "OFF SET SELF CHECK FAILURE - 1174", "INTERNAL COMM FAILURE - 1175", "INTERNAL COMM FAILURE -1474", AND "COMPRESSOR FAILURE - 1002" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions